FDA Enforcement Class II Terminated

GETINGE 46-SERIES Medical Washer-Disinfector, Models 46-4 and 46-5, with PACS 300/350 Control Series with dryer --- Class II, exempt from 510(k) requirements. For use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.

Recall: Z-0852-2015 · Reported December 31, 2014

Enforcement

Recall Number
Z-0852-2015
Event ID
69823
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Getinge Disinfection Ab
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 31, 2014
Initiation Date
November 26, 2014
Classification Date
December 23, 2014
Termination Date
May 12, 2017
Address
Ljungadalsgatan 11, Vaxjo, N/A, Sweden

Description

GETINGE 46-SERIES Medical Washer-Disinfector, Models 46-4 and 46-5, with PACS 300/350 Control Series with dryer --- Class II, exempt from 510(k) requirements. For use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.

Reason

Getinge Disinfection AB received complaints regarding sediment residuals in the manifold of the washer and disinfector of Getinge 46-Series (CM302) with PACS 300/350 control. After performing detailed root cause analysis, the firm has identified the problem as insufficient water flow throughout the manifold due to formation of air pockets where sedimentation may occur.

Code Info

Serial Numbers: SEV0810077, SEV0537108, SEV0633084, W50030454, SEV0724143, SEV0724146, SEV0609110, W50044787, SEV0533068, W50042461, W50023731, W50023732, W50038744, W50028760, SEV0802081, W50015468, W50042473, W50045852, W50048201, W50041156, SEV0911139, SEV0832021, W50031347, SEV0538174, SEV0910190, SEV0734181, W50033518, SEV0802083, W50043877, W50015494, W50047828, SEV0633086, SEV0647025, SEV0650003, SEV0650004, SEV0534069, SEV0617130, W50011492, SEV0538177, W50010176, W50027826, SEV0743108, W50012356, SEV0614021, W50039525, SEV0619121, W50021052, W50025573, W50024233, W50010768, SEV0827061, SEV0738121, SEV0804079, SEV0806117, SEV0820063, SEV0820064, W50041988, SEV0727015, SEV0739202, SEV0906031, SEV0814068, SEV0805187, W50050381, W50050415, SEV0808424, SEV0626127, SEV0645102, SEV0710062, SEV0802086, SEV0815031, SEV0815032, W50027223, W50027224, SEV0826158, W50027541, W50043237, SEV0906032, W50012254, W50038229, W50014011, W50013832, W50012253, W50036280, W50036309, W50048577, SEV0537107, SEV0842075, SEV0750143, W50016365, SEV0808427, SEV0724139, SEV0724145, W50038722, SEV0908129, SEV0908130, SEV0649007, W50023502, SEV0842076, SEV0808429, SEV0837132, SEV0827051, W50027618, SEV0705085, SEV0705084, SEV0705083, W50026375, W50023804, W50014025, SEV0742115, W50028434, W50040964, W50019155, W50010908, W50013646, W50036143, W50045596, W50029312, W50029313, SEV0806118, W50028808, W50045860, W50014993, SEV0636055, SEV0738120, W50050491, W50028245, W50035227, W50048232, SEV0626230, W50023210, W50030137, SEV0808430, SEV0724144, SEV0534064, WAA053021, W50041067, W50030350, SEV0540080, W50045855, W50032192, W50029835, W50031507, SEV0739208, SEV0802085, SEV0622093, SEV0650002, SEV0619122, W50016086, W50039500, SEV0543058, SEV0910191, SEV0837135, W50018212, SEV0537103, WAA052845, SEV0605066, W50034170, SEV0524337, W50011160, W50011491, W50032801, W50015395, W50042391, SEV0547204, W50045890, W50047809, W50013644, W50013645, W50030121, W50021208, SEV0626228, SEV0851034, SEV0827047, SEV0552041, W50037851, SEV0833107, SEV0833108, SEV0833109, SEV0833110, SEV0833111, W50036056, W50036081, SEV0837131, W50015490, SEV0716176, SEV0716177, W50043774, W50023888, SEV0738132, SEV0827050, SEV0724140, W50015641, W50045601, W50024979, SEV0614035, W50016223, W50025076, SEV0815030, SEV0820065, W50016368, SEV0647023, SEV0909043, W50037190, SEV0642143, W50017770, SEV0537110, SEV0543054, SEV0538178, W50023488, SEV0614036, SEV0614037, SEV0533067, W50029860, W50029036, SEV0814067, SEV0851035, SEV0851036, W50013647, SEV0738001, SEV0739204, SEV0810081, SEV0533069, W50016222, SEV0906030, W50014012, W50019465, W50038218, SEV0521190, W50038730, SEV0826159, W50012251, SEV0738133, SEV0619119, SEV0649010, SEV0619120, SEV0614022, W50050577, W50041166, W50048183, W50017921, SEV0739205, SEV0749123, W50042459, SEV0524338, SEV0820049, SEV0640030, W50030144, W50030181, W50028247, W50014027, W50032581, W50013666, W50042946, W50045884, W50050574, W50013831, SEV0618050, SEV0618051, SEV0727034, SEV0534074, SEV0716188, W50010079, W50033538, W50047210, W50019156, W50032618, W50015495, W50036207, SEV0733024, W50021211, SEV0827056, W50043040, W50028248, W50028468, SEV0605063, SEV0815025, SEV0824064, SEV0808426, W50036147, W50017918, W50018215, W50030353, W50036140, WAA053036, SEV0724142, SEV0533065, W50035801, SEV0907057, W50045822, SEV0836007, W50028320, W50010906, W50028809, W50023209, W50048247, W50027514, SEV0739206, SEV0739209, SEV0626231, SEV0738131, W50038686, W50011489, W50011490, W50046986, SEV0633083, W50021142, SEV0907054, W50021143, SEV0716048, W50023897, SEV0907055, W50011159, SEV0647022, SEV0851037, SEV0907056, W50021141, SEV0826157, SEV0848144, SEV0621158, W50047960, SEV0633088, SEV0705086, SEV0727019, SEV0626121, SEV0626229, W50010177, SEV0738122, SEV0738123, SEV0521184, W50016376, SEV0633087, W50019467, W50015466, W50017920, W50033104, W50025852, W50016366, SEV0632078, SEV0543059, W50048551, W50035559, SEV0521191, SEV0515157, SEV0515159, SEV0548037, SEV0644133, SEV0727032, SEV0820048, SEV0844215, SEV0906028, SEV0906029, W50010905, W50015493, W50037099, W50037204, W50039512, W50042471, W50042478, W50050571, W50050573, W50050578, WAA053175, SEV0618048, SEV0545170, SEV0724147, SEV0632079, SEV0827060, SEV0743107, SEV0820057, SEV0826160, W50029023, SEV0524334, W50011161, W50022983, W50045665, W50048002, W50012252, W50018216, W50018928, W50019154, SEV0734183, W50050575, SEV0820050, W50027512, SEV0716185, SEV0820058, W50041125, SEV0605064, SEV0615074, SEV0649008, SEV0622094, W50016364, SEV0737120, W50038421, SEV0824063, W50017919, SEV0815033, W50033839, SEV0626126, SEV0808428, W50015496, SEV0533071, W50023733, SEV0716184, SEV0742118, SEV0727033, SEV0649009, SEV0810078, W50039166, SEV0618049, W50038256, SEV0536194, SEV0716183, SEV0613001, W50042222, W50043233, SEV0636042, SEV0636043, W50039286, W50017771, SEV0804077, SEV0633085, W50021356, W50021357, SEV0532062, W50020918, W50030145, W50038244, W50036145, SEV0626120, SEV0716175, W50023501, SEV0733032, SEV0734182, SEV0738002, W50010907, W50037207, SEV0708018, W50048219, W50014030, SEV0626234, SEV0738130, W50042491, SEV0543056, SEV0536195, SEV0543055, W50013311, W50028993, SEV0739203, SEV0613098, SEV0739207, SEV0647024, W50035975, SEV0708019, SEV0702017, SEV0805191, SEV0521192, SEV0910192, W50041173, SEV0842074, W50021214, SEV0723112, W50037115, SEV0833100, SEV0833101, SEV0833102, SEV0833103, SEV0833104, W50017914, W50024126, W50029644, SEV0637017, SEV0605065, SEV0724138, W50037201, SEV0536197, W50045549, W50048239, W50011158, SEV0834133, SEV0708020, SEV0710061, SEV0849100, W50017917, W50030122, W50023198, W50018213, W50018214, SEV0533070, SEV0612048, SEV0632080, SEV0727018, SEV0537104, SEV0524333, W50024980, SEV0808423, SEV0734180, W50024130, W50024131, SEV0626125, W50035799, W50015488, SEV0827046, SEV0613100, SEV0615042, SEV0637018, SEV0543057, SEV0810080, W50021145, SEV0548036, SEV0606022, W50028246, W50028112, W50028113, W50027222, W50033877, SEV0908131, W50029035, SEV0827045, W50024981, SEV0521186, SEV0538173, SEV0827057, W50013312, SEV0626232, SEV0613099, W50016367, SEV0626233, W50018929, SEV0911140, SEV0911141, W50015398, SEV0521187, W50041062, SEV0524327, WAA052236, SEV0910189, W50021210, W50030183, SEV0804078, SEV0808425, SEV0626119, SEV0724141, W50047838, W50015491, SEV0643166, SEV0802082 --- Canadian Units: SEV0538175, SEV0613101, W50030182, W50030412.

Distribution

Distributed US (nationwide) and the country of Canada.

Quantity

551 total: USA: 547 units, Canada: 4 units