Penumbra JET 7MAX KIT Penumbra JET 7 Reperfusion Catheter+ MAX Delivery Device + Hi-Flow Aspiration Tubing REF J7XKIT132 The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Enforcement
- Recall Number
- Z-0851-2021
- Event ID
- 86989
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Penumbra Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- January 27, 2021
- Initiation Date
- December 15, 2020
- Classification Date
- January 18, 2021
- Termination Date
- May 14, 2024
- Address
- 1 Penumbra, N/A, Alameda, CA, 94502-7610, United States
Description
Penumbra JET 7MAX KIT Penumbra JET 7 Reperfusion Catheter+ MAX Delivery Device + Hi-Flow Aspiration Tubing REF J7XKIT132 The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.
Product Name: JET 7MAX (Penumbra JET 7 Xtra Flex + MAX Delivery Device) Catalog Number: J7XKIT132 UDI Code: 00815948021273 Lot Number Range: C21487 thru C21772 F97946 thru F98497
Worldwide distribution: U.S. (nationwide) including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO , MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. And O.U.S (countries) of: Canada, Chile, Colombia, Costa Rica, Dominican Republic, Hong Kong, Israel, Japan, Korea, Lebanon, Malaysia, Nepal, New Zealand, Oman, Panama, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, United Arab Emirates, and Vietnam.
15 catheters