FDA Enforcement Class II Terminated

Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy.

Recall: Z-0847-2015 · Reported December 31, 2014

Enforcement

Recall Number
Z-0847-2015
Event ID
69820
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Natus Medical Incorporated
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 31, 2014
Initiation Date
November 19, 2014
Classification Date
December 22, 2014
Termination Date
July 20, 2015
Address
5900 1st Ave S, Seattle, WA, 98108-3248, United States

Description

Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy.

Reason

Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.

Code Info

All distributed Natus neoBLUE 2 systems with LED PCB replacements received after February 16, 2012. For use only with neoBLUE devices manufactured between 2002 and 2007 (neoBLUE2)

Distribution

worldwide

Quantity

1907 total (1413 in US and 494 international)