FDA Enforcement Class II Ongoing

Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.

Recall: Z-0846-2024 · Reported February 7, 2024

Enforcement

Recall Number
Z-0846-2024
Event ID
93738
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
XTANT Medical Holdings, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 7, 2024
Initiation Date
November 28, 2023
Classification Date
January 29, 2024
Address
664 Cruiser Ln, Belgrade, MT, 59714-9719, United States

Description

Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.

Reason

Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.

Code Info

UDI-DI: 00810076564005, Lot: AP23000143

Distribution

US distribution in MI.

Quantity

6