FDA Enforcement Class II Terminated

ZYM B Reagent (REF 70493) ZYM B reagent is an additional test used to reveal the results of some miniaturized biochemical test included in the API strips. The API product line is a standardized system combining several biochemical tests, which enables group or species identifications of microorganisms.

Recall: Z-0841-2014 · Reported February 5, 2014

Enforcement

Recall Number
Z-0841-2014
Event ID
66931
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Biomerieux France Chemin De L'Or
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 5, 2014
Initiation Date
November 8, 2013
Classification Date
January 27, 2014
Termination Date
November 30, 2017
Address
Chemin De L'Orme, Marcy L'Etoile, N/A, France

Description

ZYM B Reagent (REF 70493) ZYM B reagent is an additional test used to reveal the results of some miniaturized biochemical test included in the API strips. The API product line is a standardized system combining several biochemical tests, which enables group or species identifications of microorganisms.

Reason

bioM¿rieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493); leading to a false negative results of some API biochemical tests using the ZYM B reagent.

Code Info

Lot numbers: 1002171010, 1002295110, 1002360360, 1002421440, 1002485410, 1001951820, 1002011320, and 1002071020.

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to U.S. Nationwide; Algerie, Argentina, Armenia, Australia, Austria, Azerbaidjian, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bolivia, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Chad, Chile, China, Colombia, Democratic Republic of Congo, Costa Rica, Cote d'Ivorie, Germany, Czech Republic, Dominican Republic, Dutch Antilles, Dutch Antilles, Ecuador, Egypt, Republica De El Salvador, Estonia, Germany, Ethiopia, France, Gabon, Georgia, Germany, Greece, Guatemala, Guinea, Herzegovina, Honduras, Hungary, India, Germany, Indonesia, Iraq, Israel, Italy, Jamaica, Japan, Iraq, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malaysia, Malaysia, Mali, Malta, Mexico, Romania, Mongolia, Morocco, Mozambique, Singapore, Netherlands, Managua, Niamey Niger, Nigeria, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Uganda, Kingdom of Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Syria, Tanzania, Italy, Thailand, Trinidad & Tobago, Tunisia, Turkey, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Venezuela, Vietnam, Vietnam, Yemen and Zimbabwe

Quantity

39439 kits (includes ZYM B reagents, API Listeria * API NH kits)