FDA Enforcement
Class II
Terminated
Monaco RTP System. Radiation treatment planning.
Recall: Z-0841-2013
·
Reported February 27, 2013
Enforcement
- Recall Number
- Z-0841-2013
- Event ID
- 64328
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Elekta, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 27, 2013
- Initiation Date
- February 11, 2013
- Classification Date
- February 20, 2013
- Termination Date
- January 21, 2014
- Address
- 400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227, United States
Description
Monaco RTP System. Radiation treatment planning.
Reason
The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle.
Code Info
Versions 3.20.00 and 3.20.01
Distribution
Nationwide Distribution-including the states of AK, CA, FL, NJ, NC, OH, OR, TN, TX, WA, and WI.
Quantity
13