Gendex GXPS-500 SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Enforcement
- Recall Number
- Z-0840-2020
- Event ID
- 84544
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Palodex Group Oy
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 29, 2020
- Initiation Date
- January 7, 2020
- Classification Date
- January 23, 2020
- Termination Date
- April 15, 2024
- Address
- Nahkelantie 160, Tuusula, N/A, Finland
Description
Gendex GXPS-500 SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.
Serial Number: GJ1209488, GJ1209489, GJ1209491, GJ1209492, GJ1209493, GJ1209494, GJ1209495, GJ1209496, GJ1209498, GJ1209499, GJ1209504, GJ1209505, GJ1209506, GJ1209507, GJ1209509, GJ1209510, GJ1209511, GJ1209513, GJ1209514, GJ1209516, GJ1211287, GJ1211288, GJ1211289, GJ1211290, GJ1211291, GJ1211292, GJ1211293, GJ1211294, GJ1211295, GJ1211296, GJ1211297, GJ1211298, GJ1211299, GJ1211300, GJ1211301, GJ1211302, GJ1211303, GJ1211304, GJ1211305, GJ1211307, GJ1311619, GJ1311621, GJ1311622, GJ1311623, GJ1311624, GJ1311625, GJ1311626, GJ1311627, GJ1311628, GJ1311629, GJ1311630, GJ1311631, GJ1311632, GJ1311633, GJ1311634, GJ1311635, GJ1311636, GJ1311637, GJ1311638, GJ1311639, GJ1311640, GJ1311641, GJ1311642, GJ1311643, GJ1311644, GJ1311645, GJ1311646, GJ1311647, GJ1311648, GJ1311650, GJ1311651, GJ1311653, GJ1311656, GJ1311658, GJ1311659, GJ1311660, GJ1311661, GJ1311662, GJ1311663, GJ1311664, GJ1311665, GJ1311666, GJ1311669, GJ1311670, GJ1311671, GJ1311672, GJ1311673, GJ1311674, GJ1311675, GJ1311676, GJ1311677, GJ1311678, GJ1311679, GJ1311680, GJ1311681, GJ1311682, GJ1311683, GJ1311684, GJ1311685, GJ1311686, GJ1311722 PSU's having nine-digit serial numbers beginning with four digits as follows: 1143XXXXX 1221XXXXX 1224XXXXX and 1229XXXXX
US: WI and PA OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, REPUBLIC OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM