FDA Enforcement
Class II
Terminated
Randox CALCIUM (Ca) COLORIMETRIC METHOD RX DAYTONA PLUS Cat. No. CA 8309 R1, Arsenazo Reagent 4 x 20 ml GTIN: 05055273208368
Recall: Z-0840-2019
·
Reported February 27, 2019
Enforcement
- Recall Number
- Z-0840-2019
- Event ID
- 81746
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Randox Laboratories Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- February 27, 2019
- Initiation Date
- December 4, 2018
- Classification Date
- February 15, 2019
- Termination Date
- June 18, 2020
- Address
- 55 Diamond Road, Crumlin (North), N/A, Ireland
Description
Randox CALCIUM (Ca) COLORIMETRIC METHOD RX DAYTONA PLUS Cat. No. CA 8309 R1, Arsenazo Reagent 4 x 20 ml GTIN: 05055273208368
Reason
Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.
Code Info
All lot codes
Distribution
WV, PR
Quantity
1 kit