FDA Enforcement Class II Terminated

Soredex DIGORA Optime SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

Recall: Z-0839-2020 · Reported January 29, 2020

Enforcement

Recall Number
Z-0839-2020
Event ID
84544
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Palodex Group Oy
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 29, 2020
Initiation Date
January 7, 2020
Classification Date
January 23, 2020
Termination Date
April 15, 2024
Address
Nahkelantie 160, Tuusula, N/A, Finland

Description

Soredex DIGORA Optime SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

Reason

Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.

Code Info

Serial Number: SJ1208232, SJ1208233, SJ1208234, SJ1208235, SJ1208236, SJ1208238, SJ1208248, SJ1208249, SJ1208251, SJ1208252, SJ1208253, SJ1208254, SJ1208255, SJ1208256, SJ1208257, SJ1208268, SJ1208269, SJ1208270, SJ1208296, SJ1208297, SJ1208298, SJ1208299, SJ1208300, SJ1208301, SJ1208303, SJ1208304, SJ1208305, SJ1208306, SJ1208307, SJ1208309, SJ1208409, SJ1208415, SJ1208416, SJ1208417, SJ1208418, SJ1208421, SJ1208431, SJ1208433, SJ1208434, SJ1208436, SJ1208437, SJ1208439, SJ1208440, SJ1208441, SJ1208442, SJ1208446, SJ1208447, SJ1208449, SJ1208451, SJ1208452, SJ1208453, SJ1208454, SJ1208455, SJ1208456, SJ1208457, SJ1208460, SJ1208461, SJ1208462, SJ1208465, SJ1208519, SJ1208523, SJ1208524, SJ1208533, SJ1208534, SJ1208535, SJ1208536, SJ1208537, SJ1208538, SJ1208556, SJ1208558, SJ1208559, SJ1208560, SJ1208561, SJ1208563, SJ1208571, SJ1208572, SJ1208573, SJ1208574, SJ1208575, SJ1208577, SJ1208578, SJ1208579, SJ1208580, SJ1208581, SJ1208583, SJ1208584, SJ1208585, SJ1208586, SJ1208588, SJ1208596, SJ1208597, SJ1208598, SJ1208599, SJ1208600, SJ1208601, SJ1208602, SJ1208603, SJ1208604, SJ1208605, SJ1208606, SJ1208607, SJ1208608, SJ1208609, SJ1208610, SJ1210176, SJ1210177, SJ1210178, SJ1210179, SJ1210180, SJ1210181, SJ1210184, SJ1210185, SJ1210187, SJ1210188, SJ1210189, SJ1210191, SJ1210192, SJ1210193, SJ1210194, SJ1210195, SJ1210197, SJ1210200, SJ1210201, SJ1210202, SJ1210204, SJ1210205, SJ1210206, SJ1210208, SJ1210209, SJ1210210, SJ1210213, SJ1210219, SJ1210281, SJ1210282, SJ1210283, SJ1210284, SJ1210285, SJ1210287, SJ1210288, SJ1210289, SJ1210291, SJ1210292, SJ1210293, SJ1210295, SJ1210297, SJ1210298, SJ1210300, SJ1210310, SJ1210312, SJ1210313, SJ1210314, SJ1210315, SJ1210316, SJ1210317, SJ1210318, SJ1210319, SJ1210320, SJ1210322, SJ1210323, SJ1210324, SJ1210326, SJ1210398, SJ1210400, SJ1210401, SJ1210402, SJ1210404, SJ1210416, SJ1210417, SJ1210497, SJ1210499, SJ1210500, SJ1210501, SJ1210502, SJ1210503, SJ1210504, SJ1210506, SJ1210508, SJ1210511, SJ1210512, SJ1210514, SJ1210515, SJ1210528, SJ1210529, SJ1210530, SJ1210534, SJ1210535, SJ1210536, SJ1210538, SJ1210540, SJ1210541, SJ1210542, SJ1211114, SJ1211117, SJ1211118, SJ1211121, SJ1211123, SJ1211127, SJ1211129, SJ1211130, SJ1211131, SJ1211132, SJ1211133, SJ1211134, SJ1211135, SJ1211137, SJ1211138, SJ1211224, SJ1211225, SJ1211226, SJ1211227, SJ1211228, SJ1211229, SJ1211230, SJ1211231, SJ1211232, SJ1211233, SJ1211234, SJ1211235, SJ1211236, SJ1211237, SJ1211238, SJ1211239, SJ1211240, SJ1211241, SJ1211242, SJ1211243, SJ1211244, SJ1211245, SJ1211246, SJ1211247, SJ1211248, SJ1211249, SJ1211250, SJ1211251, SJ1211252, SJ1211263, SJ1211341, SJ1211343, SJ1211345, SJ1211346, SJ1211347, SJ1211348, SJ1211349, SJ1211354, SJ1211355, SJ1211356, SJ1211358, SJ1211360, SJ1211378, SJ1211379, SJ1211382, SJ1311602, SJ1311603, SJ1311605, SJ1311606, SJ1311607, SJ1311608, SJ1311609, SJ1311610, SJ1311611, SJ1311612, SJ1311613, SJ1311614, SJ1311615, SJ1311616, SJ1311617, SJ1311766, SJ1311767, SJ1311768, SJ1311769, SJ1311770, SJ1311771, SJ1311772, SJ1311774, SJ1311776, SJ1311777, SJ1311778, SJ1311779, SJ1311780, SJ1311781, SJ1311783, SJ1312271, SJ1312481, SJ1313119, SJ1313121, SJ1313122, SJ1313124, SJ1313127, SJ1313220, SJ1313223, SJ1313224, SJ1313225, SJ1313228, SJ1313230, SJ1314362, SJ1314363, SJ1314368 PSU's having nine-digit serial numbers beginning with four digits as follows: 1143XXXXX 1221XXXXX 1224XXXXX and 1229XXXXX

Distribution

US: WI and PA OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, REPUBLIC OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM

Quantity

297