FDA Enforcement Class II Terminated

Randox CALCIUM (Ca) Colorimetric Method RX Series Cat. No. CA 3871, R1. Arsenazo Reagent 9 x 51ml GTIN: 05055273200904

Recall: Z-0839-2019 · Reported February 27, 2019

Enforcement

Recall Number
Z-0839-2019
Event ID
81746
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Randox Laboratories Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 27, 2019
Initiation Date
December 4, 2018
Classification Date
February 15, 2019
Termination Date
June 18, 2020
Address
55 Diamond Road, Crumlin (North), N/A, Ireland

Description

Randox CALCIUM (Ca) Colorimetric Method RX Series Cat. No. CA 3871, R1. Arsenazo Reagent 9 x 51ml GTIN: 05055273200904

Reason

Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.

Code Info

All lot codes

Distribution

WV, PR

Quantity

225 kits