FDA Enforcement
Class II
Terminated
DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Foot/Ankle Blanket NS, REF T656NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.
Recall: Z-0839-2014
·
Reported February 5, 2014
Enforcement
- Recall Number
- Z-0839-2014
- Event ID
- 66742
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- DeRoyal Industries Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 5, 2014
- Initiation Date
- September 17, 2013
- Classification Date
- January 27, 2014
- Termination Date
- June 5, 2017
- Address
- 1595 Highway 33 S, New Tazewell, TN, 37825-7105, United States
Description
DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Foot/Ankle Blanket NS, REF T656NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.
Reason
The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
Code Info
Lot Numbers: 32901129, 32980912, 33015541, 33024746, 33105239
Distribution
Nationwide Distribution.
Quantity
25 units