Imactis Navigation Kit, REF 10100, Content Sterile Needle Holder Ref. A10010 CE 0459, Sensor Cover Ref. PC3688 CE 0482, Adhesif Film Ref. 38.20.15 CE 0459, EN Navigation Kit. Compatible with linear instruments of diameter ranging from 0.5 mm to 3.0 mm (25G to 11G), Sterile, Rx Only,
Enforcement
- Recall Number
- Z-0838-2020
- Event ID
- 84460
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Imactis
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 29, 2020
- Initiation Date
- August 1, 2019
- Classification Date
- January 22, 2020
- Address
- 20 Rue Du Tour De L Eau, St Martin D Heres, N/A, France
Description
Imactis Navigation Kit, REF 10100, Content Sterile Needle Holder Ref. A10010 CE 0459, Sensor Cover Ref. PC3688 CE 0482, Adhesif Film Ref. 38.20.15 CE 0459, EN Navigation Kit. Compatible with linear instruments of diameter ranging from 0.5 mm to 3.0 mm (25G to 11G), Sterile, Rx Only,
The firm has become aware that the sensor cover, a component of the single-use navigation kit may contain microbial contamination. A source of microbial contamination has been identified in a raw product used to manufacture the tip of the cover.This could result in direct exposure of the intra-cardiovascular, intra-lymphatic or neurological system to a level of endotoxin sufficient to produce a pyrogenic response.
Lot # 18120001/Exp. 2020-04-01 and Lot # 19040001/Exp. 2021-01-22
US: WI & IL
122 kits