FDA Enforcement Class II Ongoing

Imactis Navigation Kit, REF 10100, Content Sterile Needle Holder Ref. A10010 CE 0459, Sensor Cover Ref. PC3688 CE 0482, Adhesif Film Ref. 38.20.15 CE 0459, EN Navigation Kit. Compatible with linear instruments of diameter ranging from 0.5 mm to 3.0 mm (25G to 11G), Sterile, Rx Only,

Recall: Z-0838-2020 · Reported January 29, 2020

Enforcement

Recall Number
Z-0838-2020
Event ID
84460
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Imactis
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 29, 2020
Initiation Date
August 1, 2019
Classification Date
January 22, 2020
Address
20 Rue Du Tour De L Eau, St Martin D Heres, N/A, France

Description

Imactis Navigation Kit, REF 10100, Content Sterile Needle Holder Ref. A10010 CE 0459, Sensor Cover Ref. PC3688 CE 0482, Adhesif Film Ref. 38.20.15 CE 0459, EN Navigation Kit. Compatible with linear instruments of diameter ranging from 0.5 mm to 3.0 mm (25G to 11G), Sterile, Rx Only,

Reason

The firm has become aware that the sensor cover, a component of the single-use navigation kit may contain microbial contamination. A source of microbial contamination has been identified in a raw product used to manufacture the tip of the cover.This could result in direct exposure of the intra-cardiovascular, intra-lymphatic or neurological system to a level of endotoxin sufficient to produce a pyrogenic response.

Code Info

Lot # 18120001/Exp. 2020-04-01 and Lot # 19040001/Exp. 2021-01-22

Distribution

US: WI & IL

Quantity

122 kits