FDA Enforcement Class II Terminated

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee Blanket w/ Straps NS, REF T654NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.

Recall: Z-0837-2014 · Reported February 5, 2014

Enforcement

Recall Number
Z-0837-2014
Event ID
66742
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
DeRoyal Industries Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 5, 2014
Initiation Date
September 17, 2013
Classification Date
January 27, 2014
Termination Date
June 5, 2017
Address
1595 Highway 33 S, New Tazewell, TN, 37825-7105, United States

Description

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee Blanket w/ Straps NS, REF T654NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.

Reason

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

Code Info

Lot Numbers: 32871775, 32980891, 33132755, 33141184, 33217572, 33329945

Distribution

Nationwide Distribution.

Quantity

48 units