FDA Enforcement
Class II
Terminated
adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
Recall: Z-0836-2015
·
Reported December 31, 2014
Enforcement
- Recall Number
- Z-0836-2015
- Event ID
- 69781
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ion Beam Applications S.A.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 31, 2014
- Initiation Date
- November 7, 2014
- Classification Date
- December 19, 2014
- Termination Date
- June 29, 2015
- Address
- Chemin du Cyclotron, 3, Louvain La Neuve, N/A, Belgium
Description
adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
Reason
Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du
Code Info
PAT.109 (US), Treatment room 4
Distribution
US Distribution in PA only.
Quantity
1