Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in China. Physical Therapy.
Recall: Z-0834-2014
·
Reported February 5, 2014
Enforcement
- Recall Number
- Z-0834-2014
- Event ID
- 66742
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- DeRoyal Industries Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 5, 2014
- Initiation Date
- September 17, 2013
- Classification Date
- January 27, 2014
- Termination Date
- June 5, 2017
- Address
- 1595 Highway 33 S, New Tazewell, TN, 37825-7105, United States
Description
Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in China. Physical Therapy.
Reason
The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
Code Info
Lot Numbers: 31480713, 31480721, 31968743, 31971619, 31990844, 32005711, 32018491, 32037764, 32118020, 32710832, 32736223, 33053264, 33105221, 33141176, 33164204, 33164407, 33276273, 33329911
Distribution
Nationwide Distribution.
Quantity
738 units