FDA Enforcement Class II Ongoing

Naturalyte Liquid Acid concentrate bottles, Product Code 08-2251-0 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Recall: Z-0833-2021 · Reported January 20, 2021

Enforcement

Recall Number
Z-0833-2021
Event ID
86939
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Fresenius Medical Care Holdings, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 20, 2021
Initiation Date
November 23, 2020
Classification Date
January 11, 2021
Address
920 Winter St Bld 950, N/A, Waltham, MA, 02451-1521, United States

Description

Naturalyte Liquid Acid concentrate bottles, Product Code 08-2251-0 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Reason

The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Code Info

Lot Numbers: 20LXAC076 20LXAC077 20LXAC091 20LXAC101 20NXAC010 Additional Affected lots as of 3/17/21: 20LXAC053 20LXAC047 20LXAC054 20LXAC049 20NXAC016 20LXAC094 20NXAC024 20NXAC025 20LXAC091 20LXAC101 20LXAC077 20LXAC048 20NXAC010 20LXAC076

Distribution

US Nationwide distribution.

Quantity

3083 cases