PTS Diagnostics CardioChek Plus Professional Analyzer, Refurbished analyzer, without box or carrying case, Reference Numbers 2702
Enforcement
- Recall Number
- Z-0833-2020
- Event ID
- 84274
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Polymer Technology Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 29, 2020
- Initiation Date
- December 2, 2019
- Classification Date
- January 21, 2020
- Termination Date
- July 6, 2020
- Address
- 7736 Zionsville Rd, Indianapolis, IN, 46268-2175, United States
Description
PTS Diagnostics CardioChek Plus Professional Analyzer, Refurbished analyzer, without box or carrying case, Reference Numbers 2702
One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.
All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00381932702010
The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.
44,448 total