Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices (e..g. modalities, workstations) Communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction that other systems have with the Centricity Enterprise archive. The intended users for Centricity Enterprise Archive are service personnel and system administrators and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-1 profiles (published by IHE).
Enforcement
- Recall Number
- Z-0833-2015
- Event ID
- 64760
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ge Healthcare It
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 31, 2014
- Initiation Date
- April 19, 2013
- Classification Date
- December 19, 2014
- Termination Date
- June 10, 2016
- Address
- 540 W Northwest Hwy, Barrington, IL, 60010-3051, United States
Description
Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices (e..g. modalities, workstations) Communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction that other systems have with the Centricity Enterprise archive. The intended users for Centricity Enterprise Archive are service personnel and system administrators and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-1 profiles (published by IHE).
GE Healthcare has become aware of a potential data loss issue associated with empty path names resulting from HL-7 updates when the cache disk is almost full.
Centricity Enterprise Archive versions: 4.0.x
Nationwide Distribution including AZ, OH, WA.
115 units