FDA Enforcement Class II Terminated

VENTANA (a member of the Roche Group) BenchMark ULTRA; REF/MU 750-600; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.

Recall: Z-0833-2013 · Reported August 14, 2013

Enforcement

Recall Number
Z-0833-2013
Event ID
64244
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ventana Medical Systems Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 14, 2013
Initiation Date
September 28, 2012
Classification Date
August 6, 2013
Termination Date
February 24, 2016
Address
1910 E Innovation Park Dr, N/A, Oro Valley, AZ, 85755-1962, United States

Description

VENTANA (a member of the Roche Group) BenchMark ULTRA; REF/MU 750-600; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.

Reason

Potiential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform. This type of overflow event can lead to migration of fluid across tubing and wiring within the instrument cabinet resulting in contact with electrical circuitry. This may create shorting conditions and u

Code Info

Cat#750-600; PN 05342716001; Serial #'s: 310724, 310735, 311187, 312055, 310373, 311052, 311171, 311953, 310750, 311266, 311267, 311564, 311565, 311596, 311304, 310481, 311924, 310607, 311627, 310527, 311436, 311742, 310132, 311432, 310034, 310408, 310687, 310697, 311343, 311958, 310728, 310778, 311991, 312070, 310471, 311326, 311952, 310103, 310777, 311446, 310924, 311303, 311528, 310490, 310913, 311529, 311494, 310139, 310281, 310465, 310275, 310617, 310665, 310716, 311853, 311871, 312137, 312101, 310240, 310237,310238, 311571, 311404, 310356, 310023, 312164, 312165, 310065, 310231, 310711, 310706, 310707, 310703, 311509, 310330, 310633, 311811, 311810, 310977, 311194, 311198, 310208, 310732, 310740, 310187, 311462, 311935, 310145, 311037, 310346, 310585, 310694, 310699, 310701, 310702, 310705, 310709, 310710, 310704, 311106, 311110, 311111, 311419, 311484, 311527, 310345, 311381, 311353, 310480, 311869, 311921, 311458, 310315, 310412, 311988, 310618, 310397, 310399, 311868, 311919, 312139, 312141, 312142, 311270, 310069, 311466, 311929, 310083, 310094, 310154, 311904, 310588, 312098, 312100, 310814, 311731, 311783, 311785, 312096, 312104, 312153, 310227, 310781, 310174, 311403, 311080, 311081, 311121, 311122, 311169, 311188, 311189, 311190, 311242, 311245, 311298, 311926, 310403, 310718, 311416, 311417, 312105, 312112, 310796, 310832, 310824, 311135, 311248, 312066, 310895, 310267, 310272, 311481, 311771, 311772, 310882, 311186, 311937, 310344, 311439, 311505, 310146, 310506, 311545, 310446, 310312, 311584, 312124, 310839, 311319, 311261, 310070, 310508, 311367, 311578, 311576, 311629, 311601, 311600, 311607, 311603, 311602, 311606, 311598, 311597, 311604, 311605, 311599, 311272, 312145, 312123, 310339, 310455, 310472, 310478, 310482, 311562, 311560, 311923, 311498, 311814, 311813, 311804, 310020, 310844, 311313, 311533, 310779, 311241, 311750, 310268, 310331, 311221, 311036, 311034, 311035, 310201, 310321, 310792, 312056, 311147, 311153, 311315, 311308, 310641, 310678, 310878, 310032, 311774, 311182, 310390, 311151, 311268, 311284, 311552, 310902, 311455, 310491, 311038, 312008, 310282, 310696, 311735, 310173, 310402, 311710, 312118, 311874, 310972, 311375, 311727, 312128, 312134, 310933, 310921, 311285, 310297, 310769, 311876, 310916, 311486, 310460, 310245, 310576, 311483, 312097, 312099, 312089, 312109, 311518, 311479, 311901, 312144, 312172, 311551, 310555, 310363, 311808, 310570, 311001, 310358, 310492, 311252, 311687, 311672, 311402, 310114, 310122, 310126, 311300, 310794, 311350, 311592, 310749, 310285, 311299, 310897, 311323, 310258, 310467, 310470, 311628, 310943, 310035, 310066, 310087, 310185, 310752, 310757, 311134, 312138, 310489, 311973, 311191, 310456, 310575, 310673, 310683, 310890, 310821, 312170, 312171, 312060, 312102, 312202, 310500, 310505, 311394, 311917, 311511, 311530, 310894, 312136, 311369, 311665, 312152, 310804, 311517, 311465, 310893, 310896, 311946, 311947, 310323, 310776, 310654, 310299, 310452, 311675, 310404, 310810, 310708, 310712, 311073, 311250, 310910, 311401, 311928, 310941, 310026, 310089, 310336, 310341, 311736, 311610, 311608, 311609, 311728, 310820, 311821, 310141, 310912, 311443, 311442, 311243, 311207, 310256, 310378, 310229, 312163, 312058, 310771, 311103, 310915, 310533, 310124, 311318, 310218, 310811, 311738, 310037, 310038, 310365, 310377, 311210, 312085, 312086, 312087, 311295, 312023, 311493, 311357, 310833, 310800, 311464, 311423, 311173, 311364, 310135, 310608, 310935, 311352, 311354, 311356, 310920, 310183, 310257, 310306, 310852, 311431, 310876, 311949, 311956, 310828, 310830, 310928, 310929, 310930, 310355, 310271, 311384, 312122, 310846, 311345, 310119, 310391, 310881, 310483, 310488, 311893, 312063, 310284, 311663, 312131, 310374, 310507, 310144, 310775, 311730, 311939, 310106, 310205, 310241, 310620, 310713, 310031, 310742, 311016, 311463, 310445, 312117, 310906, 311376, 311377, 311378, 311379, 311380, 310288, 311334, 311957, 311714, 310898, 311477, 310210, 310224, 311033, 312076, 312078, 312090, 310632, 312116, 310751, 310249, 311468, 310923, 310123, 311760, 310180, 311160, 310905, 310605, 310458, 311925, 310822, 311485, 311503, 311500, 311474, 310879, 312061, 310189, 310191, 312039, 312051, 312052, 312059, 312188, 311948, 311072, 310182, 310503, 310922, 311172, 311461, 311324, 311209, 310799, 310807, 310314, 310855, 312127, 310084, 310165, 311322, 311321, 310573, 310572, 311559, 310375, 310181, 310308, 312103, 312077, 312106, 312107, 312108, 310474, 310486, 312021, 312167, 312169, 311247, 310634, 310461, 312203, 310730, 310729, 311065, 311348, 311346, 311332, 310763, 312126, 310036, 310348, 310362, 311395, 311459, 311472, 311236, 311450, 311451, 310884, 311118, 311887, 311337, 311338, 310842, 310028, 310318, 310479, 311773, 311259, 311679, 312009, 312012, 310283, 310453, 310501, 310663, 310889, 310715, 311482, 311183, 311184, 310610, 311520, 311506, 310369, 310770, 310351, 310827, 310497, 311006, 311399, 311040, 311355, 312132, 312133, 311510, 310410, 312088, 311407, 310043, 310044, 310045, 310818, 310085, 311120, 311918, 311989, 311117, 310179, 311105, 311049, 312176, 310099, 311563, 310100, 310190, 311823, 310058, 310061, 310864, 311389, 310805, 310048, 311911, 311425, 311146, 310228, 310260, 311534, 310736, 310738, 310722, 311223, 310128, 310137, 310059, 312140, 311471, 311519, 310831, 310120, 311307, 310806, 311902, 310319, 311798, 311802, 312129, 312130, 311076, 311139, 311132, 311181, 311496, 311507, 312147, 312154, 310762, 310447, 311168, 310466, 311619, 311617, 310451, 311415, 311756, 312048, 311149, 310892, 311488, 311491, 311238, 311531, 311801, 311898, 311934, 310101, 310142, 311532, 311317, 311341, 310637, 311310, 311237, 310880, 311420, 312053, 311940, 311941, 311951, 311467, 310209, 310817, 311708, 310413, 311755, 310462, 310464, 311927, 311697, 310354, 312057, 311041, 311115, 311492, 311441, 310195, 310125, 310194, 310223, 310222, 310279, 310469, 310795, 310326, 311502, 311192, 311039, 311051, 311385, 310405, 311764, 310850, 311413, 310825, 310743, 310744, 310785, 310801, 310802, 310280, 310303, 311251, 311406, 311200, 311333, 310401, 311271, 310325, 311920, 310793, 310684, 310068, 311262, 311264, 310109, 311161, 311487, 312119, 311834, 311950, 311729, 311437, 311444, 312054, 311053, 311046, 311193, 310649, 310392, 311860, 310409, 312120, 312121, 311104, 311892, 310812, 310996, 310556, 311936, 310739, 311677, 312064, 312065, 312110, 312113, 310883, 311907, 311945, 310639, 311900, 310564, 311522, 311535, 311239, 311240, 310809, 311296, 311843, 311454, 311480, 310476, 312095, 311997, 311820, 310566, 311784, 311258, 310192, 310758, 311818, 311899, 311201, 311382, 310843, 310733, 310734, 311712, 311440, 311150, 310937, 310737, 310680, 310829, 311430, 310609, 310741, 311246, 311253, 311342, 310545, 310568, 311699, 311698, 310939, 310942, 310496, 310907, 311909, 310692, 310926, 310534, 311269, 310838, 310327, 311852, 312111, 310891, 310062, 310172, 310834, 310463, 310502, 310914, 311944, 311021, 310316, 311695, 311703, 311805, 311806, 311807, 311809, 312003,311005, 310885, 311752, 311751, 311732, 312150, 311942, 312026, 310676, 310225, 310353, 310473, 310475, 311508, 310717, 311089, 311148, 311138, 311140, 311141, 311320, 310454, 312143, 310593, 310591, 310657, 312081, 312125, 310226, 310611, 311349, 312151, 310322, 310064, 310748, 310216, 310494, 310495, 311029, 310217, 312016, 311775, 311777, 310443, 310597, 310603, 310602, 310606, 310360, 310130, 310131, 310133, 310143, 311178, 310138, 310140, 310276, 311792, 311504, 311501, 311499, 310411, 310621, 310690, 310727, 311249, 311288, 311400, 311418, 311457, 311470, 311696, 311694, 311693, 311309, 310040, 310681, 310823, 310798, 311056, 310416, 311408, 311344, 311906, 311624, 312017, 312033, 311943, 311048, 311409, 311495, 311758, 311093, 311447, 311885, 310367, 311489, 310352, 311975, 311977, 311339, 310376, 310379, 311653, 311199, 310385, 310414, 310485, 311260, 310244, 311365, 311366, 310324, 310033, 310304, 312010, 312011, 310918, 310919, 312007, 310156, 310819, 311765, 311490, 311351, 311063, 310899, 310903, 311022, 311025, 311088, 310660, 311274, 312146, 310027, 310909, 311922, 310134, 310041, 310232, 310613, 311497, 312062, 311575, 311740, 311970, 310538, 310644, 310873, 311822, 311536, 311980, 310188, 310574, 311383, 310925, 310958, 310960, 311030, 311019, 311179, 311180, 310121, 311087, 310877, 310449, 310450, 310487, 311515, 311427, 311516, 311452, 310219, 310239, 312203 All serial numbers in the market

Distribution

Worldwide distribution: USA (nationwide) and country of: Germany.

Quantity

1735 instruments