FDA Enforcement Class II Terminated

PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700

Recall: Z-0832-2020 · Reported January 29, 2020

Enforcement

Recall Number
Z-0832-2020
Event ID
84274
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Polymer Technology Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 29, 2020
Initiation Date
December 2, 2019
Classification Date
January 21, 2020
Termination Date
July 6, 2020
Address
7736 Zionsville Rd, Indianapolis, IN, 46268-2175, United States

Description

PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700

Reason

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

Code Info

All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00381932700016

Distribution

The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.

Quantity

44,448 total