HemosIL PT-Fibrinogen HS PLUS, IVD --- Instrumentation Laboratory IL Company. A high sensitivity thromboplastin reagent based on recombinant human tissue factor (RTF) for the quantitative determination in human citrated plasma of Prothrombin Time (PT) and Fibrinogen on IL Coagulation Systems. The product is used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Anticoagulant Therapy (OAT).
Enforcement
- Recall Number
- Z-0832-2015
- Event ID
- 69824
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Instrumentation Laboratory Co.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 31, 2014
- Initiation Date
- November 21, 2014
- Classification Date
- December 19, 2014
- Termination Date
- January 26, 2016
- Address
- 180 Hartwell Road, Bedford, MA, 01730-2443, United States
Description
HemosIL PT-Fibrinogen HS PLUS, IVD --- Instrumentation Laboratory IL Company. A high sensitivity thromboplastin reagent based on recombinant human tissue factor (RTF) for the quantitative determination in human citrated plasma of Prothrombin Time (PT) and Fibrinogen on IL Coagulation Systems. The product is used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Anticoagulant Therapy (OAT).
Some vials of HemosIL PT-Fibrinogen HS PLUS, Part No. 0008469810 (various lots manufactured prior to June 2014) exhibit a yellow/brownish color after reconstitution (yellow color or darker) instead of the typical white to off-white color. Vials exhibiting this yellow/brownish color were observed to cause prolonged prothrombin (PT) clotting times.
Lots: N1122175, exp. date 11/30/2014; N0132936, exp. date 1/31/2015; N0233302, exp. date 2/28/2015; N0333382, exp. date 3/31/2015; E0333382, exp. date 3/31/2015; N0333626, exp. date 3/31/2015; N0433932, exp. date 4/30/2015; N0634953, exp. date 6/30/2015; N0735317, exp. date 5/31/2015; N0735599, exp. date 7/31/ 2015; N0836106, exp. date 8/31/2015; N0936197, exp. date 9/30/2015; N0349325, exp. date 12/31/2015; N1222324, exp. date 12/31/2015; N0148094, exp. date 12/31/2015; N0248652, exp. date 2/29/2016; N0249017, exp. date 2/29/2016; N0449790, exp. date 4/30/2016; N0440009, exp. date 4/30/2016.
Worldwide Distribution: US (nationwide) including Puerto Rico; and countries of: ALBANIA, ALGERIA, ANDORRA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, BELARUS, BRAZIL, BRUNEI, BULGARIA, CHILE, CHINA, COLUMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, FRANCE, GABON, GEORGIA, GERMANY, GREECE, GUATEMALA, HONG KONG, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, KENYA, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA, MEXICO, NEW ZEALAND, PAKISTAN, PANAMA, PARAGUAY, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SERBIA, SINGAPORE, SOUTH AFRICA, SPAIN, SRI LANKA, SWITZERLAND, TURKEY, UAE, UGANDA, UKRAINE, UNITED KINGDOM, AND URUGUAY.
USA: 4088 kits; Foreign: 97,925 kits