Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX¿ Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX¿ Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX¿ Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1 4) LICOX¿ Complete Brain Probe Kit (comprised of IM1, CC1.SB) Sterile Rx Only Manufacturer CMS -Gesse;schaft for medizinische Sondentechnik mbH Dorfstrasse 2 24247 <eilkendorf Germany Distributed by Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536, USA
Enforcement
- Recall Number
- Z-0831-2015
- Event ID
- 69803
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Integra LifeSciences Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 24, 2014
- Initiation Date
- November 7, 2014
- Classification Date
- December 18, 2014
- Termination Date
- August 11, 2015
- Address
- 311 Enterprise Dr, N/A, Plainsboro, NJ, 08536-3344, United States
Description
Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX¿ Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX¿ Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX¿ Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1 4) LICOX¿ Complete Brain Probe Kit (comprised of IM1, CC1.SB) Sterile Rx Only Manufacturer CMS -Gesse;schaft for medizinische Sondentechnik mbH Dorfstrasse 2 24247 <eilkendorf Germany Distributed by Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536, USA
Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.
CC1 Licox¿ PO2 Microprobe, 200 mm 021213, 040414, 070314, 130114, 240214, 270114, 280314, 300614 CC12 Licox¿ PO2 Microprobe, 300 mm 020614, 040814, 050514, 070414, 070714, 220414 CC1.P1 Licox¿ PMO Combined Probe Containing Oxygen and Temperature 050614, 060214, 060314, 100414, 100714, 130214, 130314, 170114, 190614, 210814, 260614, 310714 IM3ST Licox¿ Complete Brain IMC-Probe Kit: 020614, 030314, 110714, 130114, 140314, 160614, 180714, 200114, 280314 IP1.P Licox¿ Single Lumen Bolt CMPS and Temperature Bolt Kit 030714, 060214, 060314, 070814, 080514, 100714, 120614, 130314, 190614, 310714 IP2.P Licox¿ Double Lumen Bolt CMPS, Temperature and ICP Monitoring Kit 030714, 050614, 060214, 060314, 070814, 080514, 090114, 100714, 130314, 140814, 200214, 200414, 260614, 290814 IT2 Licox¿ Tunneling CMPS and Temperature Monitoring Kit 090114, 100414, 190614, 200214, 260614, 310714
Nationwide, Foreign
2,506 units