FDA Enforcement Class II Terminated

Medisafe Distal Duck Kit and Duck Bag, Cleaner, ultrasonic, medical instrument - Product Usage: intended to keep instruments moist during transportation between point of use and cleaning as an initial step of reprocessing.

Recall: Z-0830-2021 · Reported January 20, 2021

Enforcement

Recall Number
Z-0830-2021
Event ID
86906
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Steris Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 20, 2021
Initiation Date
November 19, 2020
Classification Date
January 11, 2021
Termination Date
October 25, 2023
Address
5960 Heisley Rd, N/A, Mentor, OH, 44060-1834, United States

Description

Medisafe Distal Duck Kit and Duck Bag, Cleaner, ultrasonic, medical instrument - Product Usage: intended to keep instruments moist during transportation between point of use and cleaning as an initial step of reprocessing.

Reason

STERIS has identified that certain lots of diluted 4-Zyme may contain bacteria, specifically Pseudomonas fluorescens. The presence of this bacteria can cause the color of the detergent to darken over time. There is an improbable risk to users of the product from exposure to this bacteria, and no risk to patients.

Code Info

Model# M20350 Lot# 1811273,1811277,1906484,1908593,1909635,1910651, 1911658,2001730,2004770 Model# M20358 Lot# 1903408,1906508,1907556,1909644,2001718,2004772 Model# M20359 Lot# 1811278,1906499,1910649,1911680,1911691,1912708, 2001719,2004768 Model# M20400 Lot# 1901340,1901349,1902374,1904448,1904464,1905505, 1906529,1907571,1908595,1909624,1910653,1912700, 2001717,2004766,2004774,2006797

Distribution

Worldwide distribution - US Nationwide distribution including in the states of CA, DE, CO, VA, FL, WA, TX, OH, PA, MA, IA, CT and the countries of United Kingdom, Saudi Arabia, Portugal, Denmark, Ireland, United Arab Emirates, South Korea, Australia, Greece, Ashmore & Cartier Islands, Belgium, Canada, China, Germany, Japan, Norway, Poland, Qatar, Romania, Sweden, Thailand.

Quantity

23,445 units, Domestic 5,727 units, Foreign 17,718 units