Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000. Product Usage: Chromophare surgical lights are medical lights for usage in hospital treatment rooms. They provide localized lighting focused on the patient s body for detecting and treating disease, injury, and disability. The surgical lights may only be used in rooms used for medical purposes that have been equipped in accordance with VDE 0100-710 or IEC 60364-7-10 in accordance with the regulations. The Chromophare surgical lights consist of a ceiling tube, SK Box swivel arm support, and a light head. The light head is mounted on a cardanic suspension unit and can be turned, swiveled and tilted in any direction. The spatial repositioning of the light head makes use of the 360 degree rotating swivel/horizontal arm. The 360 degree rotating spring arm is used to adjust the height. The light can be mounted in several configurations, such as a single/dual ceiling mount, a boom-integrated suspension mount, a wall mount, or a mobile roll stand.
Enforcement
- Recall Number
- Z-0828-2018
- Event ID
- 79251
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Communications
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 7, 2018
- Initiation Date
- January 29, 2018
- Classification Date
- March 1, 2018
- Termination Date
- July 8, 2019
- Address
- 571 Silveron Blvd, N/A, Flower Mound, TX, 75028, United States
Description
Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000. Product Usage: Chromophare surgical lights are medical lights for usage in hospital treatment rooms. They provide localized lighting focused on the patient s body for detecting and treating disease, injury, and disability. The surgical lights may only be used in rooms used for medical purposes that have been equipped in accordance with VDE 0100-710 or IEC 60364-7-10 in accordance with the regulations. The Chromophare surgical lights consist of a ceiling tube, SK Box swivel arm support, and a light head. The light head is mounted on a cardanic suspension unit and can be turned, swiveled and tilted in any direction. The spatial repositioning of the light head makes use of the 360 degree rotating swivel/horizontal arm. The 360 degree rotating spring arm is used to adjust the height. The light can be mounted in several configurations, such as a single/dual ceiling mount, a boom-integrated suspension mount, a wall mount, or a mobile roll stand.
The device could have a missing spanner nut and snap ring which could potentially result in the equipment falling.
Serial numbers distributed in the U.S.: 022817-71019, 22817-71033, 022817-71034, 022817-71035, 022817-71048, 022817-71049, 022817-71050, 022817-71051, 030817-71104 030817-71105, 030817-71106, 030817-71107, 030817-71108, 030817-71109, 030817-71110, 030817-71111, 030817-71114, 030817-71115, 030817-71116, 030817-71117, 030817-71120, 030817-71121, 030817-71122, 030817-71123, 030817-71124, 030817-71125, 030817-71126, 030817-71127, 030817-71128, 030817-71129, 030817-71130, 030817-71131, 030817-71132, 30817-71133, 030817-71134, 030817-71135, 030817-71136, 030817-71137, 030817-71139, 030817-71140, 030817-71141, 030817-71142, 030817-71143, 030817-71144, 030817-71145, 030817-71146, 030817-71147, 030817-71148, 030817-71149, 030817-71151, 030817-71152, 030817-71153, 031417 -71180, 031417 71198, 031417-71154, 031417-71155, 031417-71156, 031417-71157, 031417-71158, 031417-71164, 031417-71165, 031417-71166, 031417-71167, 031417-71170, 031417-71171, 031417-71172, 031417-71173, 031417-71174, 031417-71175, 031417-71176, 031417-71177, 031417-71178, 031417-71181, 031417-71182, 031417-71184, 031417-71185, 031417-71186, 031417-71187, 031417-71188, 031417-71189, 031417-71190, 031417-71191, 031417-71192, 031417-71195, 031417-71196, 031417-71197, 031417-71199, 031417-71201, 031417-71202, 032017-71204, 032017-71205, 032017-71206, 032017-71207, 032017-71208, 032017-71210, 032017-71211, 032017-71212, 032017-71213, 032017-71214, 032017-71215, 032017-71216, 032017-71217, 032017-71218, 032017-71219, 032017-71220, 032017-71221, 032017-71222, 032017-71224, 032017-71225, 032017-71228, 032017-71229, 032017-71230, 032017-71231, 032017-71232, 032017-71233, 032017-71234, 032017-71235, 032017-71236, 032017-71237, 032017-71238, 032017-71239, 032017-71240, 032017-71241, 032017-71242, 032017-71243, 032017-71245, 032017-71246, 032017-71247, 032017-71248, 032017-71253, 032017-71255, 032017-71257, 032017-71258, 032017-71259, 032017-71260, 032017-71261, 032017-71262, 032017-71263, 032017-71264, 032017-71265, 032017-71266, 032017-71267, 032017-71268, 032017-71269, 032017-71270, 032017-71271, 032017-71272, 032017-71273, 032017-71274, 032017-71275, 032017-71276, 032017-71277, 032017-71278, 032017-71279, 032017-71280, 032017-71281, 032017-71282, 032017-71283, 032017-71284, 032017-71285, 032017-71286, 032017-71287, 032017-71288, 032017-71289, 032017-71290, 032017-71291, 032017-71292, 032017-71293, 032017-71295, 032017-71296, 032017-71297, 032017-71298, 032017-71299, 032017-71300, 032017-71302, 032017-71303, 033117-71304, 033117-71305, 033117-71306, 033117-71307, 033117-71308, 033117-71309, 033117-71310, 033117-71311, 033117-71312, 033117-71313, 033117-71314, 033117-71315, 033117-71316, 033117-71317, 033117-71318, 033117-71319, 033117-71320, 033117-71321, 033117-71322, 033117-71323, 033117-71324, 033117-71325, 033117-71326, 033117-71327, 033117-71328, 033117-71329, 033117-71330, 033117-71331, 033117-71332, 033117-71333, 033117-71334, 033117-71335, 033117-71336, 033117-71337, 033117-71338, 033117-71339, 033117-71340, 033117-71341, 033117-71342, 033117-71343, 033117-71344, 033117-71345, 033117-71346, 033117-71347, 033117-71348, 033117-71349, 033117-71350, 033117-71351, 033117-71352, 033117-71353, 040517-71404, 040517-71405, 040517-71406, 040517-71407, 040517-71408, 040517-71409, 040517-71410, 040517-71411, 040517-71412, 040517-71414, 040517-71415, 040517-71416, 040517-71417, 040517-71419, 040517-71420, 040517-71421, 040517-71422, 040517-71423, 040517-71426, 040517-71427, 040517-71428, 040517-71429, 040517-71430, 040517-71431, 040517-71433, 040517-71434, 040517-71435, 040517-71436, 040517-71437, 040517-71438, 040517-71439, 040517-71440, 040517-71441, 040517-71442, 040517-71444, 040517-71446, 040517-71447, 040517-71448, 040517-71449, 040517-71450, 040517-71451, 040517-71452, 040517-71453, 040517-71457, 041217-71454, 041217-71455, 041217-71456, 041217-71458, 041217-71459, 041217-71460, 041217-71461, 041217-71462, 041217-71463, 041217-71464, 041217-71465, 041217-71466, 041217-71467, 041217-71468, 041217-71469, 041217-71470, 041217-71471, 041217-71474, 041217-71475, 041217-71476, 041217-71477, 041217-71478, 041217-71479, 041217-71480, 041217-71481, 041217-71482, 041217-71483, 041217-71484, 041217-71485, 041217-71486, 041217-71487, 041217-71488, 041217-71489, 041217-71490, 041217-71491, 041217-71492, 041217-71493, 041217-71494, 041217-71495, 041217-71496, 041217-71497, 041217-71498, 041217-71499, 041217-71500, 041217-71501, 041217-71502, 041217-71503, 041817-71504, 041817-71505, 041817-71518, 041817-71519, 041817-71521, 041817-71522, 042017-71574, 042017-71575, 042017-71576, 042017-71577, 042017-71578, 042017-71579, 042017-71580, 042017-71581, 042017-71583, 042017-71584, 042017-71585, 042017-71586, 042017-71587, 042017-71588, 042017-71589, 042017-71590, 042017-71594, 042017-71595, 042017-71596, 042017-71597, 042017-71598, 042017-71599, 042017-71603, 042017-71604, 042017-71605, 042017-71606, 042017-71607, 042017-71608, 042017-71609, 042017-71610, 042017-71611, 042017-71612, 042017-71614, 042017-71615, 042017-71616, 042017-71617, 042017-71618, 042017-71619, 042017-71620, 042017-71621, 042017-71622, 042017-71623, 042017-71624, 042017-71625, 042017-71626, 042017-71627, 042017-71628, 042017-71629 042017-71630, 042017-71631, 042017-71633, 042017-71634, 042017-71635, 042017-71636, 042017-71637, 042017-71638, 042017-71639, 042017-71640, 042017-71641, 042017-71642, 042017-71643, 042017-71644, 042017-71645, 042017-71646, 042017-71647, 042017-71648, 042017-71649, 042017-71650, 042017-71651, 042017-71652, 042017-71653, 042017-71654, 042017-71655, 042017-71656, 042017-71657, 042017-71658, 042017-71660, 042017-71661, 042017-71662, 042017-71664, 042017-71667, 042017-71668, 042017-71669, 042017-71670, 042017-71671, 050317-71724, 050317-71726, 050317-71727, 050317-71728, 050317-71729, 050317-71731, 050317-71733, 050317-71734, 050317-71735, 050317-71736, 050317-71737, 050317-71738, 050317-71739, 050317-71741, 050317-71742, 050317-71743, 050317-71744, 050317-71745, 050317-71746, 050317-71747, 050317-71748, 050317-71749, 050317-71750, 050317-71751, 050317-71752, 050317-71753, 050317-71754, 050317-71755, 050317-71756, 050317-71757, 050317-71758, 050317-71759, 050317-71760, 050317-71761, 050317-71764, 050317-71765, 050317-71766, 050317-71767, 050317-71768, 050317-71769, 050317-71770, 050317-71771, 050317-71772, 050317-71773, 050817-72087, 050817-72134, 051117-71774, 051117-71775, 051117-71776, 051117-71777, 051117-71778, 051117-71780, 051117-71781, 051117-71782, 051117-71783, 051517-71788, 051517-71789, 051517-71790, 051517-71791, 051517-71792, 051517-71793, 051517-71794, 051517-71795, 051517-71796, 051517-71797, 051517-71798, 051517-71799, 051517-71800, 051517-71802, 051517-71803, 051517-71804, 051517-71805, 051517-71806, 051517-71807, 051717-71858, 051717-71859, 051717-71860, 051717-71861, 051717-71862, 051717-71863, 051717-71864, 051717-71865, 051717-71866, 051717-71868, 051717-71869, 051717-71870, 051717-71871, 051717-71872, 051717-71873, 051717-71874, 051717-71875, 051717-71876, 051717-71877, 051717-71878, 051717-71879, 051717-71880, 051717-71881, 051717-71882, 051717-71883, 051717-71884, 051717-71885, 051717-71886, 051717-71887, 051717-71888, 051717-71889, 051717-71890, 051717-71891, 051717-71892, 051717-71893, 051717-71894, 051717-71895, 051717-71896, 051717-71897, 051717-71898, 051717-71899, 051717-71900, 051717-71901, 051717-71902, 051717-71903, 051717-71904, 051717-71905, 051717-71906, 051717-71907, 051717-71908, 051717-71909, 051717-71910, 051717-71911, 051717-71912, 051717-71913, 051717-71914, 051717-71915, 051717-71916, 051717-71917, 051717-71918, 051717-71919, 051717-71921, 051717-71923, 051717-71940, 051717-71941, 051717-71942, 051717-71943, 051717-71944, 051717-71945, 051717-71946, 051717-71947, 051717-71948, 051717-71949, 051717-71950, 051717-71951, 051717-71952, 051717-71953, 051717-71954, 051717-71955, 051717-71956, 051717-71957, 053117-72021, 053117-72022, 053117-72023, 053117-72024, 053117-72025, 053117-72026, 053117-72028, 053117-72030, 053117-72031, 053117-72032, 053117-72033, 053117-72034, 053117-72035, 053117-72036, 053117-72037, 053117-72038, 053117-72039, 053117-72040, 053117-72041, 053117-72042, 053117-72043, 053117-72044, 053117-72047, 053117-72048, 053117-72049, 053117-72050, 053117-72052, 053117-72053, 053117-72054, 053117-72055, 053117-72056, 053117-72057, 053117-72058, 053117-72059, 053117-72060, 053117-72061, 053117-72062, 053117-72063, 053117-72065, 053117-72066, 053117-72067, 053117-72068, 060817-71042, 060817-72071, 060817-72072, 060817-72073, 060817-72074, 060817-72075, 060817-72076, 060817-72077, 060817-72078, 060817-72079, 060817-72080, 060817-72081, 060817-72082, 060817-72083, 060817-72084, 060817-72085, 060817-72086, 060817-72088, 060817-72089, 060817-72091, 060817-72092, 060817-72094, 060817-72095, 060817-72096, 060817-72097, 060817-72098, 060817-72099, 060817-72100, 060817-72102, 060817-72104, 060817-72105, 060817-72106, 060817-72107, 060817-72108, 060817-72109, 060817-72111, 060817-72113, 060817-72114, 060817-72115, 060817-72116, 060817-72117, 060817-72118, 060817-72119, 060817-72120, 060817-72121, 060817-72122, 060817-72123, 060817-72124, 060817-72125, 060817-72126, 060817-72127, 060817-72128, 060817-72129, 060817-72130, 060817-72131, 060817-72132, 060817-72133, 060817-72135, 060817-72136, 060817-72137, 060817-72138, 060817-72139, 060817-72140, 060817-72141, 060817-72142, 060817-72143, 060817-72144, 060817-72145, 060817-72146, 060817-72147, 060817-72148, 060817-72149, 060817-72150, 060817-72151, 060817-72152, 060817-72153, 060817-72154, 060817-72155, 060817-72156, 060817-72157, 060817-72158, 060817-72159, 060817-72161, 060817-72162, 060817-72163, 060817-72164, 060817-72165, 060817-72166, 061217-71312, 061517-72206, 061517-72207, 061517-72208, 061517-72209, 061517-72210, 061517-72211, 061517-72212, 061517-72213, 061517-72214, 061517-72215, 061517-72216, 061517-72217, 061517-72218, 061517-72219, 061517-72220, 062117-72221, 062117-72222, 062117-72223, 062117-72224, 062117-72225, 062117-72226, 062117-72227, 062117-72228, 062117-72229, 062117-72230, 062117-72231, 062117-72232, 062117-72233, 062117-72234, 062117-72235, 062317-72238, 062317-72239, 062317-72240, 062317-72242, 062317-72243, 062317-72244, 062317-72245, 062317-72246, 062317-72247, 062317-72248, 062317-72249, 062317-72250, 062617-72351, 062617-72352, 062617-72353, 062617-72354, 062617-72355, 062617-72356, 062617-72357, 062617-72358, 062617-72359, 062617-72360, 062617-72361, 062617-72362, 062617-72363, 062617-72364, 062617-72401, 062617-72402, 062617-72403, 062617-72404, 062617-72405, 062617-72406, 062617-72407, 062617-72408, 062617-72409, 062617-72410, 062617-72411, 062617-72412, 062617-72414, 062917-72416, 062917-72418, 062917-72419, 062917-72420, 062917-72422, 062917-72423, 062917-72424, 062917-72425, 062917-72426, 062917-72427, 062917-72428, 062917-72429, 062917-72430, 062917-72431, 062917-72432, 062917-72433, 062917-72434, 062917-72435, 062917-72436, 062917-72438, 062917-72439, 062917-72441, 062917-72442, 062917-72443, 062917-72445, 062917-72446, 062917-72447, 062917-72448, 062917-72449, 062917-72450, 062917-72451, 062917-72452, 062917-72453, 062917-72454, 062917-72455, 062917-72457, 062917-72460, 062917-72461, 062917-72463, 062917-72464, 062917-72464, 062917-72465, 071017-72508, 071017-72509, 071017-72510, 071017-72511, 071017-72512, 071017-72513, 071017-72514, 071017-72515, 071017-72516, 071017-72517, 071017-72518, 071017-72519, 071017-72520, 071017-72521, 071017-72522, 071117-72525, 071117-72528, 071117-72530, 071117-72531, 071117-72533, 071117-72534, 071117-72535, 071117-72536, 071117-72537, 071117-72538, 071117-72540, 071117-72541, 071117-72542, 071117-72543, 071117-72544, 071117-72546, 071117-72547, 071117-72548, 071117-72549, 071117-72550, 071117-72551, 071117-72552, 071117-72553, 071117-72554, 071117-72555, 071117-72556, 071117-72557, 071117-72558, 071117-72559, 071117-72561, 071117-72562, 071117-72563, 071117-72564, 071117-72565, 071117-72569, 071117-72570, 071117-72571, 071117-72572, 071117-72573, 071117-72575, 071117-72576, 071117-72578, 071117-72579, 071117-72580, 071117-72581, 071117-72582, 071817-72593, 071817-72594, 071817-72595, 071817-72596, 071817-72597, 071817-72598, 071817-72599, 071817-72600, 071817-72601, 071817-72602, 071817-72603, 071817-72604, 071817-72605, 071817-72607, 071817-72608, 071817-72613, 071817-72614, 071817-72615, 071817-72616, 071817-72617, 071817-72618, 071817-72619, 071817-72620, 071817-72621, 071817-72622, 071817-72623, 071817-72624, 071817-72625, 071817-72626, 071817-72627, 071817-72628, 071817-72629, 071817-72630, 071817-72631, 071817-72632, 071817-72633, 071817-72634, 071817-72636, 071817-72637, 071817-72638, 071817-72639, 071817-72640, 071817-72641, 071817-72642, 071817-72643, 071817-72644, 071817-72645, 071817-72646, 071817-72647, 071817-72648, 071817-72649, 071817-72650, 071817-72652, 071817-72851, 071917-72606, 072917-72807, 072917-72808, 072917-72809, 072917-72810, 072917-72811, 072917-72812, 072917-72814, 072917-72815, 072917-72816, 072917-72817, 072917-72820, 072917-72821, 072917-72822, 073117-72818, 073117-72859, 073117-72860, 073117-72861, 073117-72862, 073117-72863, 073117-72865, 073117-72866, 073117-72867, 073117-72868, 073117-72869, 073117-72870, 073117-72871, 073117-72872, 073117-72873, 073117-72875, 073117-72881, 080117-72884, 080117-72886, 080117-72887, 080117-72888, 080117-72889, 080117-72890, 080117-72891, 080117-72892, 080117-72893, 080117-72894, 080117-72895, 080117-72896, 080117-72897, 080117-72898, 080117-72899, 080117-72900, 080117-72901, 080117-72902, 080117-72905, 080117-72907, 080117-72909, 080117-72910, 080117-72911, 080117-72912, 080117-72913, 080117-72916, 080117-72917, 080117-72918, 080117-72919, 080117-72920, 080117-72921, 080117-72923, 080117-72924, 080417-72927, 080417-72928, 080417-72929, 080417-72930, 080417-72931, 080417-72932, 080417-72933, 080417-72934, 080417-72935, 080417-72936, 080417-72937, 080417-72938, 080417-72939, 080417-72940, 080417-72941, 080417-72942, 080417-72943, 080417-72944, 080717-72945, 080717-72946, 080717-72947, 080717-72948, 080717-72949, 080717-72950, 080717-72951, 080717-72952, 080717-72953, 080717-72955, 080717-72958, 080717-72959, 080717-72960, 080717-72961, 080717-72963, 080717-72964, 080817-72966, 080817-72967, 080817-72968, 080817-72969, 080817-72971, 080817-72973, 080817-72974, 080817-72975, 080817-72976, 080817-72980, 080817-72981, 080817-72982, 080817-72984, 080917-72987, 080917-72990, 080917-72993, 080917-72994, 080917-72995, 080917-72996, 080917-72999, 080917-73000, 080917-73002, 080917-73003, 080917-73004, 080917-73005, 080917-73006, 080917-73007, 080917-73008, 080917-73009, 080917-73016, 080917-73019, 081117-72997, 081117-73027, 081117-73029, 081117-73030, 081117-73033, 081117-73034, 081117-73035, 081117-73036, 081117-73037, 081117-73038, 081117-73039, 081117-73040, 081117-73041, 081117-73042, 081117-73043, 081117-73044, 081217-73058, 081217-73059, 081217-73060, 081217-73061, 081217-73062, 081217-73064, 081417-72926, 081417-73066, 081417-73068, 081417-73070, 081417-73072, 081417-73082, 081417-73083, 081417-73084, 081417-73085, 081417-73086, 081417-73087, 081417-73088, 081517-71788, 081617-73094, 081617-73095, 081617-73096, 081617-73097, 081617-73098, 081617-73099, 081617-73100, 081617-73101, 081617-73102, 081617-73103, 081617-73104, 081617-73105, 081617-73106, , 081617-73107, 081717-73115, 081717-73116, 081717-73117, 081717-73127, 081717-73129, 081717-73130, 081717-73131, 081817-73140, 081817-73142, 081817-73149, 081817-73150, 081817-73151, 081817-73152, 081817-73153, 081817-73154, 081817-73155, 081817-73156, and 081817-73158. Serial numbers distributed to Canada: 022817-71044, 030817-71150, 031417-71159, 031417-71160, 031417-71161, 031417-71162, 031417-71168, 032017-71226, 032017-71227, 040517-71418, 041217-71472, 041217-71473, 041817-71520, 042017-71591, 042017-71592, 042017-71663, 042017-71665, 050317-71725, 050317-71740, 050317-71762, 050817-72167, 051317-71763, 053117-72069, 053117-72070, 060817-72101, 060817-72103, 060817-72168, 060817-72169, 060817-72170, 062917-72456, 073117-72874, 073117-72876, 073117-72877, 073117-72878, 073117-72879, 073117-72880, 080917-73012, 080917-73015, 080917-73018, 080917-73021, 081617-73108, 081617-73109, 081617-73110, 081617-73112, 081617-73113, and 081717-73124
Worldwide Distribution - US Nationwide Distribution and in the countries of Canada.
1,162 devices