Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number LLCR99; c) LAMINECTOMY PACK, kit number LLLC26; d) KIT, CRANIOTOMY, kit number MMCR11J; e) KIT, NEURO SPINE, kit number MMNS20H; f) IR NEURO PACK, kit number OSIN77; g) NEURO SAH, kit number SANE32H; h) NEURO SAH, kit number SANE32I; i) S78FCH CRANI PACK (PS 132432), kit number UICR10U; j) CRANI PACK (PS039498), kit number UICR33AW; k) Laminectomy, kit number UILM77AR; l) DIAGNOSTIC NEURO IR TRAY, kit number WENE21D; m) DIAGNOSTIC NEURO IR TRAY, kit number WENE21F; n) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H; o) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H-01; p) NEURO IR TRAY, kit number WENI56F; q) NEURO IR TRAY, kit number WENI56H; r) NEURO IR TRAY, kit number WENI56H-01
Enforcement
- Recall Number
- Z-0827-2024
- Event ID
- 93656
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- American Contract Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- February 7, 2024
- Initiation Date
- November 29, 2023
- Classification Date
- January 26, 2024
- Address
- 7802 E Telecom Pkwy, Temple Terrace, FL, 33637-0928, United States
Description
Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number LLCR99; c) LAMINECTOMY PACK, kit number LLLC26; d) KIT, CRANIOTOMY, kit number MMCR11J; e) KIT, NEURO SPINE, kit number MMNS20H; f) IR NEURO PACK, kit number OSIN77; g) NEURO SAH, kit number SANE32H; h) NEURO SAH, kit number SANE32I; i) S78FCH CRANI PACK (PS 132432), kit number UICR10U; j) CRANI PACK (PS039498), kit number UICR33AW; k) Laminectomy, kit number UILM77AR; l) DIAGNOSTIC NEURO IR TRAY, kit number WENE21D; m) DIAGNOSTIC NEURO IR TRAY, kit number WENE21F; n) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H; o) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H-01; p) NEURO IR TRAY, kit number WENI56F; q) NEURO IR TRAY, kit number WENI56H; r) NEURO IR TRAY, kit number WENI56H-01
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
a) AGNE20J, UDI/DI 00191072183187, Lot Numbers: 934231, exp. 10/30/2023; 900231, exp. 4/10/2024; 990231, exp. 1/10/2024; 687221, exp. 11/9/2023; 720221, exp. 10/7/2023; 757221, exp. 8/31/2023; 870221, exp. 5/10/2023; 702221, exp. 10/25/2023; 795221, exp. 7/24/2023; b) LLCR99, UDI/DI 00191072162755, Lot Numbers: 756221, exp. 9/1/2023; 720221, exp. 10/7/2023; 654221, exp. 12/12/2023; 949231, exp. 2/20/2024; c) LLLC26, UDI/DI 00191072171900, Lot Numbers: 749221, exp. 9/8/2023; 695221, exp. 11/1/2023; 997231, exp. 1/3/2024; 920231, exp. 3/21/2024; d) MMCR11J, UDI/DI none, Lot Numbers: 744221, exp. 9/13/2023; 939231, exp. 3/2/2024; 876231, exp. 5/4/2024; e) MMNS20H, UDI/DI none, Lot Numbers: 963231, exp. 2/6/2024; 959231, exp. 2/10/2024; f) OSIN77, UDI/DI 00191072170705, Lot Numbers: 780221, exp. 8/8/2023; 728221, exp. 9/29/2023; 668221, exp. 11/28/2023; 991231, exp. 1/9/2024, 960231, exp. 2/9/2024; 939231, exp. 3/2/2024; UDI/DI 00191072190031, Lot Numbers: 912231, exp. 3/29/2024; 903231, exp. 4/7/2024; g) SANE32H, UDI/DI none, Lot Numbers: 838221, exp. 6/11/2023; 889221, exp. 4/21/2023; h) SANE32I, UDI/DI none, Lot Numbers: 673221, exp. 11/23/2023; 730221, exp. 9/27/2023; 784221, exp. 8/4/2023; 911231, exp. 3/30/2024; 969231, exp. 1/31/2024; i) UICR10U, UDI/DI none, Lot Numbers: 990231, exp. 1/10/2024; 921231, exp. 3/20/2024; 885231, exp. 4/25/2024; j) UICR33AW, UDI/DI none, Lot Numbers: 658221, exp. 12/8/2023; 931231, exp. 3/10/2024; 876231, exp. 5/4/2024; k) UILM77AR, UDI/DI 00191072139184, Lot Numbers: 953231, exp. 2/16/2024; 905231, exp. 4/5/2024; l) WENE21D, UDI/DI 00191072167958, Lot Numbers: 839221, exp. 6/10/2023; m) WENE21F, UDI/DI 00191072167958, Lot Numbers: 819221, exp. 6/30/2023; n) WENE21H, UDI/DI 00191072179326, Lot Numbers: 729221, exp. 9/28/2023; 686221, exp. 11/10/2023; o) WENE21H-01, UDI/DI 00191072179326, Lot Numbers: 980231, exp. 1/20/2024; p) WENI56F, UDI/DI 00191072169341, Lot Numbers: 799221, exp. 7/20/2023; q) WENI56H, UDI/DI 00191072179333, Lot Numbers: 700221, exp. 10/27/2023; r) WENI56H-01, UDI/DI 00191072179333, Lot Numbers: 652221, exp. 12/14/2023; 996231, exp. 1/4/2024
US Nationwide distribution.
2717 units