FDA Enforcement
Class II
Ongoing
Medicrea Pass LP Trocar K-wire 01.6mm x 500mm-Nitinol Product reference : A06010130
Recall: Z-0827-2020
·
Reported January 29, 2020
Enforcement
- Recall Number
- Z-0827-2020
- Event ID
- 84595
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medicrea International
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- January 29, 2020
- Initiation Date
- September 16, 2019
- Classification Date
- January 18, 2020
- Address
- 5389 Route De Strasbourg, Vancia, Rillieux La Pape, N/A, France
Description
Medicrea Pass LP Trocar K-wire 01.6mm x 500mm-Nitinol Product reference : A06010130
Reason
Conical Cannulated Screws may not accept 1.6mm K-Wire, it cannot enter the minimum tolerance axis threaded
Code Info
Lot number(s) : 18G0037
Distribution
Nationwide
Quantity
30 units