Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.
Recall: Z-0827-2016
·
Reported March 2, 2016
Enforcement
- Recall Number
- Z-0827-2016
- Event ID
- 73241
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Draeger Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 2, 2016
- Initiation Date
- February 4, 2016
- Classification Date
- February 25, 2016
- Termination Date
- April 1, 2020
- Address
- 3135 Quarry Rd, N/A, Telford, PA, 18969-1042, United States
Description
Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.
Reason
The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.
Code Info
Catalog #s 8416400 and 8417400
Distribution
US Nationwide Distribution including Puerto Rico.
Quantity
2553