FDA Enforcement Class I Terminated

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Recall: Z-0827-2016 · Reported March 2, 2016

Enforcement

Recall Number
Z-0827-2016
Event ID
73241
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Draeger Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 2, 2016
Initiation Date
February 4, 2016
Classification Date
February 25, 2016
Termination Date
April 1, 2020
Address
3135 Quarry Rd, N/A, Telford, PA, 18969-1042, United States

Description

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Reason

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

Code Info

Catalog #s 8416400 and 8417400

Distribution

US Nationwide Distribution including Puerto Rico.

Quantity

2553