FDA Enforcement Class I Ongoing

VALITUDE CRT-P, EL (Model Number U125) and EL MRI (Model Number U128)

Recall: Z-0826-2025 · Reported January 22, 2025

Enforcement

Recall Number
Z-0826-2025
Event ID
95969
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Boston Scientific Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 22, 2025
Initiation Date
December 12, 2024
Classification Date
January 10, 2025
Address
4100 Hamline Ave N, Saint Paul, MN, 55112-5700, United States

Description

VALITUDE CRT-P, EL (Model Number U125) and EL MRI (Model Number U128)

Reason

A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.

Code Info

U125 - UDI-DI: 00802526559389, 00802526559396, 00802526577109, 00802526578793, 00802526577024, Serial Numbers: 100319 100323 100329 100331 100333 100379 100380 100384 100385 100386 100390 100392 100439 100448 100452 100453 100462 100466 100467 100480 100481 100482 100485 100486 100495 100496 100497 100498 100499 100500 100502 100505 100506 100510 100514 100515 100517 100518 100524 100526 100539 100540 100541 100546 100548 100549 100552 100585 100650 100653 100656 100657 100658 100690 100692 100693 100697 100699 100700 100701 100703 100707 100711 100716 100721 100722 100723 100725 100726 100730 100731 100733 100735 100740 100743 100746 100748 100751 100754 100755 100756 100757 100758 100759 100760 100761 100763 100764 100767 100768 100770 100772 100774 100775 100776 100777 100778 100779 100780 100781 100784 100786 100788 100789 100790 100792 100793 100799 100801 100802 100803 100805 100813 100814 100818 100823 100830 100833 100843 100846 100854 100864 100866 100870 100871 100873 100877 100884 100886 100891 100910 100912 700042 700066 700191 700272 700277 700306 700507 700582 700630 700713 700726 700735 700762 700796 700801 700802 700805 700820 700832 700834 700837 700839 700852 700853 700854 700856 700860 700863 700867 700985 701002 701003 701008 701015 701023 701033 701060 701104 701156 701168 701232 701485 701520 701560 701562 701563 701564 701566 701569 701570 701571 701573 701574 701579 701580 701586 701587 701593 701599 701600 701601 701603 701604 701606 701609 701612 701613 701616 701617 701618 701620 701623 701624 701625 701626 701645 701646 701647 701648 701649 701650 701651 701653 701655 701656 701657 701659 701662 701675 701676 701678 701684 701691 701692 701693 701696 701697 701699 701700 701701 701702 701703 701704 701705 701707 701709 701710 701713 701718 701720 701722 701724 701725 701726 701729 701733 701739 701741 701742 701745 701746 701750 701751 701752 701754 701755 701757 701761 701763 701764 701765 701766 701767 701770 701771 701772 701774 701780 701781 701789 701790 701791 701795 701798 701816 701818 701823 701827 701828 701829 701833 701837 701841 701848 701856 701858 701869 701873 701879 701887 701889 701895 701896 701910 701916 701921 701928 701950 701954 701966 701981 701983 701991 701992 701999 702000 702002 702004 702012 702024 702029 702046 702047 702056 702064 702065 702066 702097 702099 702157 702235 702237 702242 702256 702260 702288 702328 702331 702359 702360 702361 702369 702379 702381 702384 702427 702431 702433 702436 702438 702439 702441 702445 702455 702458 702466 702467 702469 702470 702473 702474 702475 702476 702478 702482 702494 702498 702505 702506 702512 702515 702516 702517 702518 702529 702542 702553 702569 702574 702587 702592 702594 702596 702602 702616 702619 702620 702627 702628 702629 702630 702635 702638 702640 702644 702649 702652 702659 702663 702665 702674 702686 702687 702691 702692 702695 702702 702707 702710 702724 702731 702732 702734 702737 702743 702744 702746 702748 702751 702758 702768 702774 702803 702804 702808 702813 702814 702820 702839 702851 702863 702880 702881 702889 702922 702925 702929 702943 702953 702957 702968 702976 702977 702985 702996 703000 703002 703004 703012 703013 703014 703017 703020 703037 703039 703047 703050 703052 703062 703065 703067 703069 703096 703101 703102 703103 703112 703116 703117 703118 703120 703121 703128 703129 703131 703133 703135 703138 703141 703142 703144 703146 703147 703150 703152 703154 703156 703157 703160 703163 703167 703176 703189 703193 703195 703200 703204 703211 703214 703215 703222 703223 703225 703227 703229 703234 703236 703240 703248 703249 703250 703252 703253 703254 703256 703257 703267 703268 703281 703282 703284 703288 703290 703293 703297 703301 703304 703305 703313 703315 703323 703326 703331 703338 703342 703344 703348 703351 703356 703361 703364 703366 703369 703371 703374 703375 703376 703379 703383 703386 703387 703390 703405 703410 703414 703417 703418 703420 703428 703440 703449 703455 703457 703459 703460 703478 703479 703487 703488 703490 703505 703523 703524 703532 703544 703553 703562 703598 703604 703613 703616 703617 703639 703649 703653 703656 703657 703661 703671 703675 703677 703681 703687 703691 703692 703701 703702 703709 703714 703715 703717 703720 703724 703726 703727 703733 703734 703738 703744 703745 703746 703751 703752 703760 703763 703764 703766 703767 703768 703769 703771 703773 703775 703776 703780 703781 703785 703786 703789 703796 703804 703809 703814 703822 703825 703878 703884 703892 703896 703901 703914 703918 703921 703922 703924 703931 703937 703942 703943 703944 703946 703950 703952 703956 703957 703959 703967 703972 703978 703989 703990 703991 703992 703993 703995 703996 703999 704000 704001 704002 704005 704006 704007 704010 704014 704020 704021 704024 704026 704028 704029 704035 704041 704042 704044 704045 704047 704050 704052 704054 704055 704056 704058 704059 704064 704069 704072 704074 704076 704077 704078 704080 704085 704089 704091 704095 704096 704097 704208 704210 704246 704248 704253 704255 704257 704260 704262 704278 704299 704300 704301 704304 704307 704309 704311 704312 704314 704315 704321 704323 704340 704342 704344 704351 704354 704402 704425 704438 704441 704444 704445 704447 704449 704456 704482 704483 704484 704545 704602 704604 704615 704616 704801 704803 704820 704822 704826 704828 704829 704931 704932 704935 704936 704952 704953 705001 705002 705004 705005 705016 705018 705020 705021 705027 705029 705071 705097 705107 705108 705113 705115 705116 705117 705121 705132 705136 705148 705190 705193 705197 705221 705226 705227 705239 705248 705270 705273 705277 705288 705291 705292 705301 705320 705328 705446 705492 705502 705504 705510 705528 705535 705548 705554 705570 705575 705576 705580 705581 705582 705587 705588 705589 705594 705595 705596 705599 705605 705606 705631 705642 705646 705647 705648 705662 705663 705666 705667 705685 705688 705692 705712 705753 705759 705762 705764 705768 705783 705786 705796 705803 705808 705820 705821 705856 705860 705868 705884 705886 705888 705889 705890 705893 705894 705896 705898 705899 705901 705902 705903 705904 705905 705936 705942 705943 705949 705953 705954 706014 706016 706022 706027 706042 706048 706051 706052 706054 706055 706056 706059 706075 706076 706083 706088 706096 706099 706120 706173 706174 706189 706193 706197 706198 706211 706212 706214 706217 706232 706243 706248 706250 706251 706254 706255 706261 706273 706283 706288 706310 706314 706334 706354 706355 706362 706363 706367 706372 706377 706380 706383 706384 706388 706407 706411 706416 706418 706419 706421 706422 706424 706429 706438 706442 706444 706445 706453 706454 706455 706457 706477 706479 706480 706483 706485 706490 706501 706508 706525 706529 706551 706553 706555 706562 706564 706576 706578 706587 706588 706589 706597 706612 706614 706615 706620 706648 706691 706700 707249 707250 707263 707265 707269 707276 707277 707303 707308 707309 707313 707321 707322 707330 707341 707351 707352 707353 707355 707358 707359 707360 707361 707363 707364 707369 707504 707526 707530 707531 707532 707535 707537 707541 707552 707555 707571 707623 707624 707625 707626 707628 707638 707643 707668 707770 707781 707806 100322 100324 100326 100334 100336 100338 100391 100451 100507 100509 100511 100512 100520 100521 100709 100739 100741 100796 100797 100815 100816 100820 100831 100836 100838 100839 100848 100855 100857 100859 100862 100863 100867 100872 100882 100885 100898 100899 100903 100911 100914 100918 100920 100927 100937 700001 700004 700011 700014 700020 700026 700031 700034 700036 700037 700038 700041 700049 700051 700053 700054 700055 700057 700058 700061 700062 700065 700068 700069 700070 700074 700076 700079 700086 700087 700090 700091 700093 700095 700097 700098 700101 700107 700111 700112 700113 700114 700119 700123 700127 700129 700131 700133 700135 700136 700139 700154 700161 700162 700164 700166 700167 700168 700169 700171 700172 700173 700174 700175 700176 700177 700178 700179 700181 700182 700183 700184 700185 700187 700190 700192 700193 700194 700195 700231 700232 700233 700240 700247 700248 700257 700258 700259 700260 700263 700266 700267 700269 700275 700276 700282 700283 700287 700289 700290 700300 700312 700369 700402 700404 700436 700518 700521 700589 700675 700700 700702 700746 700747 700759 700771 700786 700806 700816 700823 700838 700843 700848 700850 700859 700875 700879 701208 701517 701572 701589 701594 701621 701627 701630 701632 701686 701711 701748 701769 701784 701799 701804 701806 701808 701817 701847 701850 701851 701854 701857 701876 701880 701881 701883 701897 701902 701904 701906 701912 701919 701922 701927 701932 701943 701951 701956 701963 701965 701967 701969 701985 701995 701997 702003 702006 702009 702010 702013 702014 702016 702017 702022 702026 702027 702039 702042 702049 702068 702069 702131 702134 702219 702238 702266 702293 702299 702300 702305 702333 702335 702338 702345 702363 702375 702420 702422 702426 702440 702447 702451 702452 702460 702479 702480 702487 702490 702491 702492 702493 702496 702497 702499 702500 702503 702504 702519 702532 702533 702538 702550 702551 702554 702563 702571 702572 702575 702577 702583 702585 702589 702598 702599 702605 702608 702611 702613 702614 702625 702633 702642 702647 702670 702675 702677 702681 702683 702684 702697 702739 702750 702761 702766 702767 702777 702785 702787 702788 702798 702801 702802 702821 702827 702830 702833 702856 702866 702872 702879 702931 702942 702947 702954 702960 702964 702966 702970 702974 702978 702984 702987 702993 702999 703001 703005 703008 703009 703010 703022 703025 703027 703029 703030 703034 703054 703097 703099 703109 703113 703140 703165 703168 703171 703172 703175 703183 703184 703199 703208 703209 703213 703219 703220 703228 703231 703233 703239 703241 703243 703260 703261 703274 703327 703328 703352 703355 703357 703362 703367 703380 703381 703384 703385 703392 703399 703400 703404 703406 703409 703411 703423 703424 703425 703435 703441 703443 703446 703447 703451 703453 703458 703463 703468 703470 703474 703491 703492 703497 703506 703521 703531 703533 703546 703549 703595 703680 703683 703684 703706 703718 703728 703736 703742 703748 703753 703782 703787 703791 703792 703793 703797 703798 703799 703800 703801 703802 703803 703805 703806 703807 703811 703813 703818 703819 703820 703821 703824 703828 703829 703832 703851 703853 703856 703857 703870 703877 703887 703889 703893 703898 703905 703906 703915 703919 703920 703934 703940 703941 703947 703948 703962 703963 704003 704009 704017 704018 704023 704027 704030 704032 704039 704043 704060 704088 704090 704239 704268 704269 704298 704302 704305 704306 704326 704332 704337 704339 704341 704345 704355 704365 704437 704439 704442 704443 704448 704480 704481 704485 704531 704536 704572 704768 704769 704816 704821 704824 704840 704844 704849 704933 704934 704958 705022 705028 705073 705074 705088 705089 705092 705093 705100 705109 705112 705114 705118 705122 705124 705126 705128 705130 705134 705137 705155 705156 705159 705171 705173 705174 705175 705180 705187 705191 705210 705212 705230 705241 705268 705274 705280 705302 705313 705490 705494 705501 705509 705526 705537 705538 705540 705543 705547 705550 705551 705552 705553 705558 705559 705560 705562 705563 705564 705567 705569 705571 705572 705573 705574 705583 705604 705608 705617 705640 705641 705649 705650 705653 705654 705655 705658 705659 705661 705676 705684 705691 705694 705696 705699 705702 705707 705708 705709 705716 705718 705719 705728 705769 705773 705781 705784 705789 705793 705800 705806 705811 705816 705825 705861 705862 705867 705873 705880 705885 705907 705926 705930 705933 705937 705940 705944 705947 706001 706002 706006 706010 706019 706020 706021 706023 706026 706028 706029 706093 706121 706125 706128 706129 706171 706175 706220 706221 706229 706230 706231 706236 706237 706259 706278 706281 706286 706291 706293 706294 706295 706301 706323 706324 706327 706328 706331 706332 706333 706335 706336 706337 706339 706349 706352 706353 706361 706373 706376 706420 706458 706482 706503 706552 706565 706567 706570 706577 706581 706596 706600 706616 706619 707076 707077 707091 707092 707093 707095 707099 707254 707258 707272 707275 707279 707288 707301 707306 707314 707315 707317 707318 707332 707343 707345 707368 707370 707508 707527 707529 707536 707556 707557 707573 707575 707582 707612 707627 707631 707636 707637 707639 707640 707641 707648 707742 707743 707771 707777 707798 707801; U128 - UDI-DI: 00802526559402, 00802526559419, 00802526576522, 00802526578120, 00802526577031, 00802526593284, Serial Numbers: 100719 100721 100724 100725 100726 100727 100730 100731 100732 100735 100736 100737 100740 100742 100744 100748 100749 100750 100753 100757 100759 100761 100763 100765 100768 100769 100772 100773 100774 100777 100784 100785 100787 100791 100793 100794 100798 100801 100805 100807 100808 100811 100816 100820 100821 100822 100825 100829 100831 100833 100834 100835 100839 100841 100842 100849 100850 100852 100853 100854 100855 100856 100859 100861 100865 100867 100870 100871 100872 100874 100880 100881 100882 100885 100886 100888 100895 100898 100903 100905 100907 100921 100925 100939 100940 100941 100943 100945 100947 100949 100952 100953 100954 100955 100956 100957 100959 100961 100965 100967 100971 100972 100974 100975 100979 100980 100982 100983 100985 100987 100988 100990 100991 100992 100994 100995 100998 101001 101007 101012 101014 101018 101021 101022 101024 101026 101027 101039 101041 101044 101052 101054 101057 101058 101066 101069 101072 101075 101077 101079 101080 101081 101087 101088 101089 101090 101092 101102 101103 101104 101107 101109 101110 101115 101117 101118 101139 101140 101143 101146 101147 101189 101193 101194 101195 101196 101199 101210 101211 101212 101213 101227 101230 101233 101235 101238 101240 101242 101244 101247 101248 101250 101251 101261 101266 101267 101269 101282 101283 101285 101289 101290 101303 101304 101306 101341 101343 101374 101375 101377 101387 101395 101398 101404 101410 101412 101415 101442 101445 101484 101493 101687 101921 101934 101950 101954 101965 101966 101968 101977 102033 102035 102036 102037 102039 102079 102082 102085 102089 102092 102093 102097 102099 102102 102103 102106 102126 102128 102132 102144 102147 102191 102263 102301 102312 102315 102327 102371 102437 102462 102469 102470 102472 102473 700065 700091 700105 700169 700222 700242 700244 700245 700252 700256 700306 700308 700311 700312 700313 700318 700323 700324 700325 700327 700328 700329 700330 700332 700339 700340 700341 700343 700346 700350 700354 700355 700356 700358 700359 700360 700364 700374 700378 700380 700388 700415 700418 700421 700424 700434 700436 700439 700441 700445 700452 700453 700456 700458 700459 700486 700490 700491 700501 700504 700532 700547 700549 700555 700561 700566 700568 700570 700572 700573 700574 700579 700580 700594 700596 700597 700598 700600 700601 700602 700603 700605 700607 700608 700609 700610 700618 700619 700620 700623 700625 700627 700630 700636 700638 700640 700641 700642 700647 700648 700649 700651 700652 700658 700659 700666 700669 700672 700677 700678 700679 700680 700682 700687 700690 700691 700694 700699 700700 700702 700705 700706 700707 700708 700713 700714 700716 700717 700719 700720 700723 700800 700801 700806 700808 700814 700816 700817 700818 700820 700822 700824 700830 700831 700833 700835 700839 700840 700897 700898 700901 700904 700906 700909 700916 700918 700920 700921 700924 700935 700936 700971 700972 700976 700977 700978 700980 700983 700984 700987 700988 700992 700994 700995 700996 700998 701000 701001 701005 701006 701007 701011 701013 701023 701026 701028 701029 701032 701035 701039 701041 701042 701046 701049 701050 701052 701054 701057 701058 701059 701081 701083 701084 701087 701088 701094 701095 701097 701099 701101 701102 701105 701106 701109 701111 701112 701117 701119 701124 701127 701136 701141 701146 701147 701149 701151 701152 701153 701154 701155 701156 701158 701166 701169 701182 701183 701188 701189 701190 701193 701194 701195 701196 701206 701207 701209 701210 701211 701214 701218 701223 701224 701226 701228 701230 701247 701249 701251 701256 701259 701263 701264 701267 701271 701272 701294 701295 701296 701302 701313 701314 701315 701317 701333 701335 701337 701339 701341 701342 701343 701344 701345 701347 701349 701350 701352 701355 701361 701363 701364 701373 701376 701383 701387 701388 701390 701391 701392 701393 701394 701395 701398 701401 701404 701405 701406 701409 701415 701418 701423 701430 701433 701434 701435 701439 701448 701449 701450 701455 701458 701459 701462 701467 701474 701480 701481 701482 701486 701490 701495 701497 701499 701501 701505 701507 701508 701513 701514 701518 701521 701522 701524 701525 701526 701530 701537 701541 701547 701551 701552 701556 701557 701561 701562 701564 701574 701576 701585 701586 701588 701591 701595 701596 701600 701602 701606 701615 701621 701623 701627 701632 701634 701635 701636 701637 701641 701642 701644 701646 701648 701651 701652 701654 701668 701669 701678 701680 701681 701686 701694 701695 701696 701697 701700 701702 701703 701708 701712 701714 701727 701728 701732 701736 701739 701740 701741 701744 701745 701755 701763 701766 701781 701782 701786 701789 701792 701796 701811 701815 701816 701817 701819 701821 701825 701827 701828 701829 701832 701834 701838 701842 701846 701847 701892 701898 701899 701901 701907 701920 701921 701925 701930 701931 701934 701941 701945 701952 701953 701954 701957 701963 701964 701965 701968 701975 701986 702033 702036 702085 702087 702089 702090 702131 702132 702145 702148 702152 702153 702154 702159 702161 702166 702167 702179 702182 702186 702187 702198 702199 702200 702204 702213 702218 702219 702225 702227 702228 702231 702233 702234 702237 702238 702239 702240 702241 702246 702248 702252 702253 702255 702256 702257 702261 702265 702266 702269 702270 702273 702274 702275 702276 702278 702281 702283 702284 702285 702286 702288 702291 702292 702293 702294 702295 702296 702297 702298 702301 702302 702309 702320 702321 702326 702328 702347 702349 702352 702355 702356 702360 702362 702363 702364 702368 702371 702372 702377 702382 702383 702389 702410 702414 702419 702421 702427 702428 702515 702522 702523 702525 702528 702572 702576 702579 702582 702584 702587 702588 702589 702590 702598 702601 702605 702608 702611 702612 702613 702614 702616 702620 702621 702626 702633 702640 702653 702659 702677 702680 702684 702685 702691 702693 702701 702721 702728 702731 702733 702737 702739 702747 702748 702751 702767 702768 702772 702775 702777 702779 702781 702784 702789 702790 702792 702796 702798 702800 702801 702804 702809 702812 702817 702818 702822 702825 702828 702829 702831 702832 702837 702838 702846 702848 702855 702860 702862 702864 702866 702871 702874 702876 702878 702879 702881 702882 702884 702885 702886 702887 702889 702892 702893 702894 702897 702898 702899 702900 702903 702904 702906 702907 702908 702910 702912 702913 702919 702922 702925 702927 702928 702936 702937 702944 702946 702947 702950 702955 702956 702957 702958 702960 702962 702971 702974 702976 702984 702986 702989 702991 702993 702994 702995 702996 702997 703001 703002 703005 703006 703010 703011 703012 703013 703014 703015 703024 703025 703026 703028 703032 703033 703034 703035 703038 703039 703042 703044 703046 703048 703050 703051 703054 703059 703060 703061 703062 703064 703065 703068 703070 703072 703075 703076 703079 703082 703084 703085 703099 703105 703108 703110 703114 703115 703116 703119 703122 703124 703126 703128 703130 703131 703136 703138 703139 703140 703150 703151 703157 703161 703162 703164 703170 703176 703182 703184 703185 703186 703198 703215 703221 703245 703256 703262 703274 703291 703294 703297 703328 703340 703352 703357 703372 703374 703378 703379 703400 703404 703405 703412 703418 703419 703422 703423 703425 703426 703437 703442 703443 703444 703445 703446 703451 703452 703453 703454 703457 703459 703461 703465 703468 703473 703477 703478 703479 703480 703482 703484 703493 703494 703503 703505 703507 703510 703511 703512 703514 703515 703517 703521 703522 703524 703527 703529 703533 703537 703538 703539 703540 703542 703543 703544 703546 703548 703552 703553 703561 703562 703563 703567 703570 703571 703572 703577 703580 703587 703588 703590 703591 703593 703594 703595 703597 703601 703602 703603 703604 703606 703618 703619 703622 703623 703634 703637 703648 703649 703650 703651 703656 703658 703660 703665 703666 703667 703670 703671 703673 703677 703679 703680 703685 703688 703689 703697 703706 703708 703712 703719 703720 703726 703731 703733 703737 703741 703742 703745 703746 703752 703753 703755 703761 703767 703770 703772 703777 703779 703783 703784 703785 703792 703793 703799 703800 703801 703808 703810 703811 703814 703819 703823 703826 703827 703828 703829 703831 703832 703834 703835 703838 703844 703847 703851 703852 703853 703862 703865 703866 703874 703878 703880 703897 703904 703909 703923 703925 703928 703933 703937 703938 703946 703947 703951 703953 703955 703956 703961 703965 703966 703967 703976 703977 703999 704000 704002 704003 704005 704008 704013 704026 704028 704031 704033 704034 704053 704059 704061 704066 704071 704072 704074 704081 704086 704097 704099 704111 704112 704116 704119 704122 704131 704139 704141 704148 704160 704173 704180 704184 704188 704189 704191 704199 704201 704203 704206 704207 704208 704210 704213 704214 704215 704216 704223 704224 704225 704226 704228 704229 704230 704231 704232 704233 704235 704236 704239 704240 704241 704255 704256 704257 704258 704261 704269 704279 704280 704283 704284 704285 704288 704290 704335 704336 704352 704354 704356 704357 704361 704362 704363 704369 704370 704393 704394 704399 704400 704401 704404 704409 704410 704422 704435 704436 704437 704438 704591 704607 704610 704614 704621 704624 704627 704628 704629 704632 704638 704642 704643 704646 704649 704651 704653 704654 704657 704658 704661 704662 704665 704668 704673 704677 704679 704682 704684 704685 704688 704692 704693 704694 704695 704698 704699 704700 704701 704702 704706 704707 704710 704714 704715 704721 704726 704731 704732 704747 704748 704749 704753 704754 704757 704758 704761 704762 704763 704766 704773 704774 704776 704783 704785 704794 704795 704798 704799 704818 704821 704822 704823 704825 704826 704827 704829 704831 704832 704844 704854 704857 704859 704866 704867 704871 704875 704877 704880 704881 704882 704883 704884 704885 704888 704891 704913 704936 704938 704941 704946 704949 704953 704957 704958 704959 704964 704971 704976 704977 704978 704979 704980 704981 704984 704990 704991 704992 704994 704996 704997 705001 705005 705006 705017 705020 705045 705049 705052 705086 705087 705444 705445 705610 705611 705614 705615 706117 706120 706127 706128 706220 706225 706227 706228 706234 706236 706240 706594 706605 706609 706611 706612 706613 706618 706621 706627 706633 706637 706644 706656 706660 706664 706666 706672 706673 706679 706682 706687 706699 706702 706715 706721 706723 706724 706725 706732 706736 706739 706745 706747 706757 706759 706789 706793 706797 706802 706804 706805 706829 706833 706834 706841 706858 706861 706862 706863 706878 706881 706882 706883 706884 706885 706889 706890 706891 706894 706895 706896 706899 706908 706915 706917 706918 706920 706921 706927 706928 706929 706930 706931 706933 706935 706961 706967 706968 706970 706976 706977 706986 706991 706992 707001 707004 707005 707011 707014 707019 707035 707046 707052 707058 707065 707071 707074 707079 707080 707083 707084 707085 707086 707087 707088 707098 707099 707100 707102 707105 707108 707110 707113 707115 707121 707122 707124 707126 707129 707132 707135 707136 707138 707139 707145 707148 707150 707156 707161 707164 707165 707167 707169 707172 707174 707177 707221 707222 707224 707225 707226 707230 707231 707233 707237 707238 707248 707250 707258 707267 707270 707287 707288 707291 707292 707296 707299 707301 707304 707305 707308 707309 707310 707312 707322 707324 707325 707327 707333 707334 707341 707342 707344 707348 707354 707360 707371 707376 707377 707381 707383 707384 707387 707389 707394 707395 707396 707407 707412 707426 707430 707431 707437 707439 707440 707441 707443 707446 707448 707451 707457 707458 707461 707463 707468 707471 707479 707480 707487 707492 707500 707522 707524 707525 707527 707531 707535 707538 707539 707546 707548 707552 707553 707554 707624 707626 707627 707630 707632 707633 707635 707636 707638 707640 707643 707644 707648 707656 707688 707692 707693 707696 707697 707698 707701 707702 707706 707707 707708 707710 707711 707714 707715 707720 707724 707725 707726 707727 707728 707733 707734 707736 707738 707744 707746 707749 707750 707751 707752 707753 707754 707755 707757 707761 707767 707771 707774 707778 707779 707781 707782 707784 707785 707787 707790 707794 707797 707798 707799 707802 707803 707804 707805 707807 707808 707812 707816 707818 707819 707820 707822 707824 707831 707832 707835 707836 707839 707840 707846 707847 707848 707849 707852 707853 707860 707861 707862 707864 707867 707868 707877 707878 707881 707889 707891 707894 707953 707959 707961 707967 707971 707993 707995 707997 707999 708023 708026 708027 708031 708041 708056 708061 708064 708065 708066 708067 708068 708069 708070 708074 708075 708078 708079 708089 708102 708104 708105 708108 708109 708110 708111 708112 708115 708118 708132 708136 708166 708190 708196 708198 708199 708202 708203 708204 708205 708210 708211 708221 708314 708317 708338 708349 708526 708601 708604 708606 708609 708611 708629 708633 708638 708642 708643 708656 708658 708659 708660 708662 708675 708681 708683 708687 708688 708689 708696 708698 708700 708709 708711 708714 708715 708717 708761 708770 708771 708773 708777 708790 708802 708805 708809 708812 708821 708825 708831 708836 708837 708845 708850 708851 708857 708860 708861 708862 708863 708868 708871 708872 708880 708884 708885 708887 708891 708895 708897 708900 708901 708902 708903 708914 708917 708918 708921 708925 708926 708928 708930 708933 708935 708936 708937 708940 708942 708947 708949 708956 708957 708963 708964 708970 708972 708975 708978 708979 708983 708985 708998 709006 709009 709034 709051 709052 709055 709058 709061 709062 709074 709093 709096 709105 709106 709110 709111 709112 709114 709122 709126 709127 709132 709133 709134 709137 709138 709141 709142 709145 709147 709148 709149 709152 709153 709155 709156 709159 709163 709164 709165 709167 709174 709189 709190 709191 709195 709208 709229 709234 709235 709236 709240 709242 709243 709245 709257 709258 709259 709263 709269 709273 709275 709280 709283 709286 709289 709294 709296 709297 709299 709302 709307 709308 709309 709312 709320 709324 709326 709330 709333 709334 709338 709342 709343 709344 709348 709351 709352 709357 709359 709365 709367 709370 709375 709376 709377 709381 709382 709389 709392 709394 709401 709402 709403 709407 709408 709409 709411 709417 709419 709420 709421 709422 709423 709424 709426 709427 709429 709432 709435 709438 709441 709442 709445 709446 709449 709451 709456 709457 709460 709463 709465 709466 709467 709469 709477 709478 709480 709484 709485 709486 709487 709491 709492 709500 709505 709506 709507 709571 709575 709576 709579 709588 709590 709599 709606 709619 709636 709660 709672 709689 709690 709705 709718 709740 709747 709753 709754 709755 709756 709769 709773 709777 709780 709785 709798 709801 709802 709810 709811 709840 709843 709858 709879 709889 709890 709928 710002 710026 710031 710038 710087 710094 710151 710183 710195 710197 710198 710199 710206 710207 710210 710212 710214 710232 710234 710242 710246 710253 710261 710263 710264 710274 710275 710277 710286 710290 710292 710300 710302 710304 710309 710310 710313 710314 710316 710318 710320 710325 710329 710332 710333 710336 710337 710340 710341 710345 710351 710355 710357 710359 710360 710364 710371 710373 710374 710379 710384 710390 710391 710393 710396 710397 710399 710400 710401 710402 710404 710405 710411 710419 710436 710437 710442 710444 710446 710448 710458 710461 710462 710468 710469 710471 710474 710480 710481 710482 710486 710489 710490 710492 710493 710536 710537 710538 710539 710553 710554 710555 710556 710573 710576 710581 710584 710587 710592 710761 710790 710792 710834 710865 710866 710897 710904 710907 710913 710939 710946 710984 710989 710994 710997 711001 711007 711013 711014 711026 711044 711046 711050 711058 711069 711074 711076 711082 711083 711087 711091 711093 711094 711095 711096 711102 711105 711107 711109 711113 711117 711118 711120 711121 711145 711147 711151 711155 711171 711172 711174 711177 711179 711183 711187 711191 711192 711196 711197 711204 711207 711213 711214 711215 711217 711220 711221 711230 711233 711235 711238 711241 711243 711245 711250 711259 711265 711273 711278 711281 711282 711284 711290 711297 711299 711304 711305 711313 711316 711317 711320 711322 711326 711330 711331 711336 711337 711341 711346 711354 711362 711368 711375 711376 711378 711380 711381 711386 711387 711394 711395 711400 711402 711403 711404 711405 711406 711410 711413 711416 711418 711424 711427 711429 711431 711439 711443 711444 711449 711450 711452 711455 711458 711463 711464 711466 711471 711475 711476 711477 711478 711480 711484 711490 711491 711501 711504 711505 711510 711511 711513 711514 711516 711521 711537 711546 711549 711556 711557 711563 711564 711565 711568 711571 711574 711577 711579 711589 711590 711593 711594 711605 711610 711613 711615 711617 711620 711623 711641 711643 711648 711653 711660 711661 711665 711670 711671 711672 711675 711676 711677 711679 711682 711684 711690 711701 711704 711708 711717 711719 711720 711724 711727 711728 711731 711733 711736 711743 711748 711753 711756 711758 711764 711769 711773 711775 711777 711780 711781 711785 711786 711787 711788 711790 711824 711827 711831 711832 711833 711836 711844 711846 711850 711851 711865 711877 711879 711882 711885 711886 711889 711894 711897 711898 711905 711907 711909 711924 711929 711941 711943 711945 711946 711950 711962 711969 711972 711984 711985 711986 711987 711996 711998 712000 712001 712004 712005 712007 712011 712016 712019 712021 712022 712024 712028 712030 712031 712035 712037 712042 712043 712047 712048 712049 712051 712053 712054 712058 712059 712060 712068 712071 712073 712074 712075 712076 712077 712081 712083 712090 712091 712094 712095 712097 712100 712102 712103 712104 712107 712109 712110 712111 712115 712116 712117 712122 712127 712130 712131 712132 712137 712139 712141 712142 712143 712145 712147 712148 712149 712152 712155 712157 712160 712164 712166 712169 712172 712175 712178 712182 712185 712191 712195 712196 712198 712201 712202 712205 712210 712211 712215 712220 712227 712229 712230 712233 712236 712239 712243 712244 712246 712248 712252 712253 712254 712257 712258 712260 712262 712268 712271 712274 712275 712277 712278 712279 712282 712284 712287 712288 712290 712291 712292 712305 712306 712312 712314 712319 712324 712330 712335 712336 712344 712353 712361 712364 712368 712372 712376 712383 712384 712386 712389 712390 712392 712396 712397 712398 712400 712401 712406 712409 712411 712412 712418 712419 712420 712421 712422 712425 712428 712429 712430 712435 712438 712439 712442 712443 712446 712447 712449 712451 712452 712471 712475 712476 712478 712479 712488 712489 712490 712492 712494 712495 712496 712497 712498 712499 712500 712502 712503 712504 712505 712506 712508 712510 712513 712517 712524 712534 712538 712541 712546 712547 712548 712551 712553 712555 712563 712566 712568 712571 712573 712576 712578 712579 712583 712585 712586 712587 712593 712594 712598 712601 712602 712605 712611 712613 712614 712617 712618 712619 712624 712626 712629 712630 712632 712635 712638 712639 712650 712661 712662 712664 712674 712676 712677 712678 712682 712685 712689 712690 712693 712694 712695 712696 712697 712702 712705 712708 712715 712718 712719 712720 712721 712722 712725 712727 712734 712737 712739 712740 712744 712747 712750 712751 712754 712755 712756 712760 712761 712766 712768 712770 712775 712777 712778 712779 712782 712783 712786 712790 712792 712793 712794 712797 712799 712800 712801 712803 712804 712805 712807 712811 712812 712815 712816 712820 712821 712822 712823 712824 712825 712826 712827 712828 712829 712832 712833 712834 712835 712836 712840 712844 712847 712850 712854 712855 712860 712861 712863 712864 712865 712869 712870 712871 712875 712879 712882 712883 712889 712893 712895 71289

More Code Info

8 712901 712907 712910 712911 712913 712917 712925 712931 712932 712933 712938 712939 712944 712953 712958 712960 712961 712962 712963 712965 712967 712968 712969 712971 712972 712974 712975 712977 712978 712979 712980 712981 712982 712983 712985 712986 712988 712989 712990 712993 712995 712996 712997 712998 712999 713000 713001 713002 713006 713007 713008 713010 713013 713014 713017 713018 713019 713020 713021 713022 713023 713025 713028 713029 713032 713034 713037 713038 713040 713041 713042 713044 713045 713047 713048 713049 713052 713054 713056 713057 713058 713059 713060 713062 713064 713068 713069 713075 713078 713082 713083 713084 713085 713086 713087 713089 713093 713095 713101 713104 713109 713110 713111 713113 713119 713120 713121 713123 713128 713129 713131 713132 713133 713134 713135 713139 713143 713146 713148 713150 713157 713162 713165 713166 713167 713168 713170 713178 713179 713181 713182 713187 713190 713191 713192 713193 713195 713199 713200 713203 713205 713207 713211 713212 713213 713216 713221 713222 713223 713225 713227 713228 713230 713231 713237 713244 713245 713249 713250 713251 713257 713259 713260 713266 713271 713272 713280 713282 713283 713284 713298 713304 713305 713308 713314 713316 713321 713333 713334 713344 713357 713359 713361 713369 713380 713385 713386 713392 713397 713402 713412 713413 713422 713466 713478 713480 713483 713485 713496 713501 713506 713507 713508 713509 713511 713546 713549 713552 713601 713602 713624 713626 713627 713636 713637 713647 713653 713655 713661 713662 713664 713667 713668 713670 713671 713673 713691 713693 713694 713695 713697 713700 713704 713705 713706 713707 713713 713719 713720 713723 713726 713727 713729 713736 713738 713746 713758 713763 713795 713799 713810 713818 713830 713835 713847 713849 713853 713856 713865 713878 713889 713898 713903 713915 713916 713920 713923 713927 713929 713932 713933 713935 713938 713939 713944 713949 713954 713955 713956 713964 713965 713966 713968 713969 713970 713971 713974 713976 713977 713978 713986 713998 713999 714007 714042 714045 714046 714048 714050 714058 714059 714061 714063 714067 714069 714070 714077 714078 714080 714082 714084 714085 714088 714095 714098 714099 714101 714111 714112 714116 714117 714122 714123 714126 714129 714130 714131 714132 714135 714152 714154 714156 714157 714159 714160 714165 714166 714172 714173 714185 714187 714188 714205 714208 714209 714210 714212 714216 714222 714225 714226 714228 714234 714239 714242 714248 714249 714258 714270 714271 714273 714276 714278 714285 714287 714291 714292 714293 714297 714298 714299 714300 714303 714304 714306 714308 714309 714311 714312 714314 714316 714319 714321 714323 714325 714326 714327 714330 714335 714338 714339 714342 714343 714345 714346 714347 714348 714349 714350 714355 714356 714361 714362 714365 714366 714367 714368 714369 714370 714371 714374 714376 714378 714382 714385 714390 714391 714393 714395 714396 714397 714398 714401 714402 714408 714409 714424 714441 714442 714456 714457 714458 714466 714469 714481 714484 714486 714488 714490 714494 714495 714497 714499 714502 714503 714504 714505 714506 714508 714511 714512 714515 714516 714517 714519 714520 714521 714523 714529 714531 714533 714534 714535 714542 714543 714577 714580 714583 714586 714587 714588 714589 714590 714591 714592 714593 714594 714596 714597 714598 714600 714612 714614 714615 714616 714619 714621 714624 714625 714629 714631 714642 714643 714649 714653 714658 714664 714667 714669 714671 714675 714678 714679 714681 714683 714687 714689 714695 714698 714699 714703 714704 714707 714708 714709 714712 714714 714718 714722 714723 714724 714728 714730 714740 714750 714752 714753 714758 714761 714767 714769 714772 714774 714780 714788 714800 714802 714803 714805 714810 714814 714820 714821 714828 714829 714832 714833 714834 714835 714837 714839 714840 714841 714842 714844 714846 714847 714848 714850 714851 714852 714853 714854 714855 714856 714858 714863 714864 714865 714866 714867 714868 714869 714870 714873 714874 714875 714876 714877 714880 714883 714884 714886 715106 715213 715216 715227 715378 715386 715387 715396 715399 715401 715402 715404 715405 715406 715407 715408 715411 715413 715416 715419 715423 715433 715436 715438 715441 715444 715446 715447 715452 715453 715454 715455 715456 715457 715458 715459 715460 715462 715465 715468 715469 715472 715474 715475 715478 715479 715485 715487 715490 715491 715493 715497 715498 715499 715500 715501 715502 715504 715506 715508 715509 715510 715511 715512 715515 715520 715522 715523 715525 715526 715530 715542 715550 715551 715554 715555 715562 715563 715564 715567 715570 715574 715576 715580 715581 715587 715591 715595 715599 715601 715607 715611 715613 715614 715615 715617 715618 715620 715623 715625 715626 715627 715629 715630 715643 715655 715664 715665 715667 715669 715670 715678 715751 715771 715784 715811 715817 715829 715841 715842 715845 715846 715849 715851 715855 715860 715866 715871 715872 715875 715876 715878 715880 715881 715888 715896 715898 715902 715903 715931 715941 715946 715947 715969 715970 715971 715974 715978 715980 715985 715988 715990 716005 716009 716029 716043 716045 716046 716047 716051 716052 716053 716071 716079 716080 716081 716085 716087 716092 716093 716095 716096 716099 716103 716104 716105 716111 716112 716113 716117 716120 716132 716144 716147 716148 716149 716154 716155 716166 716167 716168 716169 716174 716177 716179 716183 716187 716188 716204 716205 716212 716213 716214 716216 716219 716220 716223 716225 716227 716236 716237 716243 716246 716247 716249 716250 716261 716265 716266 716267 716269 716273 716275 716329 716330 716376 716377 716380 716401 716418 716441 716445 716446 716447 716448 716449 716460 716471 716474 716477 716481 716483 716485 716486 716488 716493 716499 716502 716505 716509 716513 716518 716572 716573 716575 716623 716624 716626 716629 716630 716636 716637 716641 716645 716646 716647 716648 716656 716661 716662 716669 716672 716674 716675 716679 716681 716683 716684 716691 716693 716696 716698 716699 716702 716709 716710 716712 716715 716718 716721 716724 716725 716726 716727 716729 716730 716737 716738 716742 716743 716746 716751 716752 716758 716759 716760 716863 716879 717006 717008 717009 717010 717012 717013 717016 717017 717018 717019 717024 717041 717046 717047 717049 717050 717131 717133 717135 717139 717268 717270 717273 717277 717279 717295 717304 717310 717319 717331 717333 717340 717341 717342 717344 717346 717348 717354 717370 717371 717377 717380 717387 717389 717390 717391 717394 717401 717406 717411 717412 717416 717418 717421 717422 717424 717429 717449 717470 717472 717487 717488 717491 717493 717507 717509 717512 717513 717515 717516 717518 717519 717544 717549 717551 717560 717570 717620 717622 717623 717647 717648 717654 717655 717658 717680 717702 717704 717705 717716 717720 717771 717777 717778 717779 717780 717782 717783 717789 717792 717799 717801 717802 717803 717805 717808 717812 717813 717816 717823 717826 717828 717835 717837 717838 717839 717840 717841 717845 717849 717851 717852 717862 717873 717877 717881 717882 717892 717893 717908 717909 717911 717912 717913 717914 717927 717933 717934 717935 717936 717942 717951 717968 717974 717979 717992 717995 718014 718018 718032 718034 718035 718038 718043 718066 718075 718076 718081 718087 718088 718101 718103 718106 718151 718153 718222 718225 718226 718227 718231 718232 718234 718235 718236 718238 718239 718248 718255 718256 718258 718261 718262 718268 718269 718270 718271 718276 718278 718280 718281 718282 718285 718286 718290 718293 718294 718296 718298 718301 718303 718309 718311 718312 718314 718315 718317 718318 718319 718320 718322 718323 718324 718325 718329 718330 718333 718339 718340 718341 718342 718343 718345 718346 718347 718348 718349 718350 718357 718359 718363 718365 718367 718369 718370 718374 718376 718377 718378 718379 718384 718387 718389 718390 718392 718396 718397 718398 718399 718401 718402 718404 718405 718407 718413 718414 718415 718419 718421 718424 718429 718431 718432 718476 718479 718482 718483 718485 718486 718488 718500 718502 718516 718526 718531 718535 718536 718548 718552 718553 718560 718563 718565 718568 718571 718574 718578 718580 718582 718584 718585 718588 718589 718590 718592 718594 718595 718597 718600 718601 718605 718611 718613 718615 718616 718617 718618 718620 718621 718636 718637 718639 718643 718648 718649 718652 718661 718663 718665 718666 718683 718689 718694 718698 718701 718710 718715 718718 718719 718725 718728 718735 718750 718761 718763 718782 718785 718788 718789 718795 718811 718821 718824 718828 718829 718830 718840 718845 718849 718857 718862 718866 718869 718874 718876 718878 718893 718904 718909 718910 718912 718914 718921 718923 718925 718961 718963 718964 718969 718970 718971 718974 718977 718981 718982 718983 718986 718988 718990 718996 718997 719001 719005 719006 719014 719015 719020 719021 719026 719028 719030 719032 719033 719036 719037 719038 719042 719043 719045 719051 719053 719054 719055 719056 719057 719058 719059 719062 719064 719067 719069 719070 719072 719073 719074 719076 719077 719079 719089 719090 719093 719096 719105 719110 719114 719115 719116 719117 719121 719127 719128 719129 719138 719146 719187 719188 719236 719440 719491 719493 719495 719516 719649 719650 719654 719665 719671 719673 719674 719675 719676 719677 719678 719679 719743 719747 719748 719751 719756 719761 719763 719765 719766 719768 719772 719775 719777 719780 719783 719784 719785 719786 719787 719789 719795 719803 719805 719806 719809 719811 719812 719817 719818 719825 719829 719830 719831 719833 719835 719836 719837 719838 719842 719844 719847 719848 719849 719850 719851 719852 719853 719857 719864 719866 719867 719870 719872 719873 719874 719876 719879 719880 719882 719891 719893 719895 719897 719898 719899 719900 719903 719905 719907 719908 719943 719955 719957 719962 719964 719976 719978 719981 719985 719987 719989 720000 720007 720011 720013 720014 720019 720027 720029 720043 720046 720062 720063 720065 720069 720070 720081 720082 720086 720087 720088 720090 720092 720097 720098 720099 720102 720103 720104 720108 720112 720113 720118 720127 720134 720143 720144 720145 720146 720150 720155 720157 720162 720163 720165 720167 720168 720169 720170 720176 720177 720179 720194 720236 720291 720298 720299 720301 720303 720312 720313 720316 720324 720326 720328 720329 720335 720347 720349 720350 720352 720354 720366 720368 720377 720384 720396 720400 720402 720404 720416 720417 720418 720424 720426 720427 720428 720429 720431 720432 720433 720434 720435 720437 720440 720444 720445 720447 720448 720449 720451 720457 720458 720462 720464 720465 720466 720470 720471 720473 720475 720476 720478 720480 720483 720486 720490 720492 720493 720494 720495 720498 720501 720505 720506 720508 720510 720512 720513 720518 720519 720520 720521 720529 720532 720534 720539 720541 720542 720545 720547 720548 720550 720553 720554 720556 720557 720559 720571 720572 720578 720596 720598 720599 720603 720604 720611 720615 720623 720648 720657 720658 720660 720666 720675 720678 720681 720712 720718 720735 720738 720739 720743 720744 720745 720753 720766 720768 720774 720776 720777 720779 720780 720781 720785 720799 720803 720805 720809 720816 720817 720819 720831 720832 720833 720834 720840 720843 720846 720847 720850 720852 720853 720857 720868 720875 720876 720881 720886 720896 720897 720901 720902 720903 720920 720936 720938 720940 720941 720946 720947 720948 720954 720955 720956 720959 720960 720961 720964 720965 720985 720986 720987 721016 721017 721021 721024 721031 721036 721040 721041 721043 721044 721049 721050 721053 721055 721056 721057 721059 721062 721063 721065 721066 721068 721071 721072 721073 721078 721079 721082 721084 721085 721086 721087 721088 721091 721092 721094 721095 721096 721108 721109 721113 721114 721115 721119 721121 721123 721125 721126 721127 721128 721129 721131 721132 721134 721136 721138 721139 721140 721142 721143 721144 721147 721148 721152 721155 721156 721157 721158 721159 721162 721163 721164 721165 721168 721169 721170 721172 721174 721200 721203 721240 721241 721348 721350 721351 721352 721353 721550 721557 721559 721563 721566 721573 721574 721583 721585 721588 721589 721591 721592 721595 721598 721599 721600 721605 721610 721627 721630 721643 721647 721648 721653 721654 721655 721666 721667 721671 721673 721674 721677 721679 721680 721682 721683 721685 721699 721704 721710 721712 721715 721724 721726 721735 721736 721741 721742 721747 721756 721763 721766 721767 721769 721771 721773 721774 721786 721787 721788 721789 721791 721793 721794 721797 721798 721801 721805 721806 721817 721821 721822 721823 721832 721833 721834 721836 721840 721841 721845 721847 721860 721861 721862 721864 721866 721867 721874 721877 721883 721886 721896 721897 721902 721904 721908 721910 721911 721915 721926 721934 721936 721938 721941 721945 721949 721955 721961 721962 721963 721965 721967 721968 721969 721972 721974 721975 721977 721978 722040 722041 722042 722046 722060 722071 722072 722073 722081 722084 722095 722096 722098 722112 722115 722117 722118 722174 722187 722199 722201 722233 722239 722242 722244 722245 722255 722259 722260 722295 722319 722322 722327 722333 722335 722337 722339 722340 722347 722349 722353 722356 722358 722359 722360 722369 722372 722374 722379 722380 722381 722382 722385 722386 722388 722389 722390 722393 722396 722397 722398 722400 722401 722402 722404 722405 722409 722420 722423 722424 722425 722427 722429 722435 722439 722446 722452 722456 722460 722465 722467 722468 722472 722473 722476 722477 722478 722480 722482 722484 722490 722492 722493 722499 722500 722501 722503 722505 722508 722517 722527 722528 722530 722531 722539 722541 722545 722550 722551 722552 722553 722554 722556 722558 722559 722565 722568 722613 722633 722660 722683 722685 722688 722689 722691 722695 722698 722699 722700 722701 722723 722724 722729 722734 722736 722740 722755 722756 722757 722758 722759 722761 722762 722763 722764 722765 722766 722767 722770 722771 722775 722777 722778 722779 722780 722785 722787 722792 722793 722794 722795 722799 722802 722805 722806 722807 722810 722812 722813 722814 722815 722817 722819 722821 722822 722823 722825 722826 722827 722831 722834 722835 722838 722839 722840 722842 722844 722845 722849 722852 722856 722857 722858 722860 722861 722862 722863 722867 722870 722871 722873 722874 722876 722877 722878 722881 722884 722886 722887 722888 722890 722891 722899 722902 722903 722905 722909 722912 722914 722915 722917 722918 722919 722922 722925 722926 722928 722930 722931 722932 722933 722935 722942 722944 722946 722947 722950 722952 722955 722958 722960 722963 722964 722965 722966 722968 722969 722972 722975 722977 722978 722979 722981 722985 722992 722999 723000 723002 723003 723004 723005 723006 723007 723010 723012 723013 723017 723020 723025 723027 723030 723031 723032 723036 723037 723041 723042 723046 723048 723049 723052 723053 723059 723060 723062 723063 723064 723067 723068 723070 723073 723076 723083 723085 723090 723093 723096 723097 723099 723102 723103 723107 723110 723111 723112 723114 723115 723118 723120 723122 723125 723127 723128 723130 723132 723138 723140 723141 723142 723143 723146 723149 723154 723155 723156 723158 723160 723161 723164 723168 723169 723171 723172 723174 723175 723176 723179 723183 723184 723185 723186 723189 723195 723197 723206 723209 723210 723218 723219 723220 723221 723222 723223 723274 723275 723276 723282 723291 723295 723299 723414 723415 723418 723420 723443 723446 723452 723519 723520 723523 723525 723527 723528 723529 723530 723534 723537 723538 723539 723540 723542 723553 723554 723572 723585 723587 723597 723598 723599 723602 723612 723626 723628 723631 723633 723654 723656 723657 723658 723659 723660 723662 723664 723665 723669 723670 723671 723677 723678 723683 723684 723685 723691 723714 723718 723734 723735 723750 723764 723765 723774 723775 723776 723777 723778 723779 723780 723814 723819 723821 723822 723834 723837 723838 723842 723846 723849 723850 723854 723856 723864 723866 723867 723870 723871 723881 723890 723962 723994 723995 724011 724016 724082 724085 724105 724198 724285 724416 725255 725261 725265 725917 725931 725932 725933 725969 725971 725972 725976 725979 725981 725984 725985 725986 725987 725994 726000 726001 726010 726011 726014 726026 726035 726038 726040 726042 726048 726049 726050 726052 726054 726076 726077 726093 726095 726118 726129 726131 726134 726137 726139 726185 726186 726192 726218 726221 726222 726227 726228 726230 726233 726238 726239 726243 726246 726247 726281 726282 726283 726287 726308 726312 726323 726324 726327 726329 726332 726334 726360 726581 726589 726590 726618 726623 726624 726625 726626 726629 726633 726634 726667 726693 726698 726700 726702 726704 726705 726707 726711 726714 726722 726724 726725 726729 726736 726739 726740 726751 726752 726760 726762 726763 726764 726765 726791 726798 726799 726800 726801 726811 726812 726813 726814 726815 726817 726828 726829 726830 726831 726861 726871 726880 726886 726895 726897 726898 726903 726905 726906 726944 726945 726946 726950 726955 726959 726963 726987 726991 727039 727063 727080 727081 727086 727096 727098 727103 727104 727106 727107 727132 727133 727138 727139 727140 727142 727169 727177 727179 727184 727186 727187 727190 727191 727199 727200 727201 727203 727204 727208 727229 727230 727231 727232 727235 727265 727266 727267 727270 727280 727282 727284 727285 727286 727288 727299 727302 727303 727315 727318 727335 727338 727344 727345 727352 727357 727392 727398 727401 727402 727405 727432 727434 727436 727449 727453 727463 727481 727506 727562 727571 727572 727591 727605 727621 727625 727650 727654 727670 727672 727854 727858 727863 727867 727868 727975 728521 728522 708062 708094 708865 100733 100899 100948 100999 101032 101042 101101 101446 102087 102131 102135 102175 102255 102261 102311 102314 102316 102320 102337 102349 102350 102352 102353 102354 102359 102363 102366 102367 102398 102400 102401 102403 102404 102407 102434 102435 102436 102459 102461 102463 102467 102468 102471 102475 102480 102481 700136 700144 700219 700227 700230 700253 700254 700275 700285 700372 700429 700443 700449 700455 700466 700507 700519 700676 700681 700682 700828 700841 700986 701016 701157 701227 701260 701407 701422 701472 701483 701503 701516 701540 701575 701589 701611 701614 701620 701629 701638 701639 701640 701643 701645 701655 701683 701684 701707 701710 701715 701716 701723 701966 702050 702052 702083 702092 702136 702144 702156 702177 702214 702229 702412 702422 702425 702448 702450 702473 702516 702530 702595 702606 702638 702648 702663 702671 702672 702686 702694 702698 702706 702717 702760 702778 702802 702807 702810 702813 702833 702857 702921 702939 702945 702948 702949 702954 702967 702968 702969 703053 703058 703063 703077 703092 703094 703112 703165 703173 703196 703202 703205 703229 703249 703250 703252 703265 703313 703339 703488 703535 703928 704044 704163 704170 704211 704212 704218 704243 704268 704278 704281 704286 704383 704384 704385 704397 704408 704423 704427 704429 704439 704440 704441 704611 704647 704652 704670 704839 704843 704850 704852 704893 704974 704988 704995 705022 705616 706602 706729 706816 706909 707072 707101 707120 707141 707162 707168 707173 707223 707269 707293 707345 707347 707358 707363 707373 707424 707429 707469 707472 707699 707748 707823 707825 707828 707874 707892 707893 708000 708029 708058 708095 708117 708122 708130 708151 708208 708271 708339 708344 708438 708599 708605 708618 708649 708650 708670 708671 708674 708677 708685 708706 708707 708823 708833 708834 708839 708841 708847 708852 708866 708869 708870 708873 708882 708883 708888 708896 708904 708906 708910 708912 708915 708916 708944 708948 708951 708959 709103 709104 709121 709128 709129 709135 709170 709175 709192 709201 709237 709238 709276 709277 709278 709314 709345 709366 709410 709413 709488 709570 709577 709586 709630 709741 709866 709959 710108 710121 710122 710200 710223 710237 710255 710256 710259 710271 710273 710281 710284 710288 710323 710330 710331 710338 710346 710356 710361 710363 710365 710377 710378 710383 710385 710388 710389 710395 710407 710408 710412 710414 710418 710420 710423 710438 710439 710445 710450 710452 710456 710459 710466 710470 710473 710476 710485 710487 710488 710552 710769 710889 710943 710944 710950 710979 710991 711009 711015 711018 711039 711054 711081 711089 711090 711092 711097 711106 711111 711116 711122 711123 711140 711153 711154 711157 711218 711222 711251 711254 711256 711275 711292 711312 711327 711328 711358 711360 711365 711366 711370 711374 711388 711393 711414 711415 711419 711423 711440 711495 711498 711506 711509 711550 711578 711583 711716 711738 711752 711825 711925 711932 711947 711949 711961 711997 712029 712036 712041 712046 712084 712088 712105 712112 712121 712134 712144 712150 712151 712153 712162 712167 712171 712176 712180 712261 712265 712280 712298 712303 712371 712427 712481 712493 712511 712512 712521 712535 712570 712625 712634 712637 712655 712658 712698 712699 712710 712711 712716 712742 712745 712746 712759 712774 712787 712853 712892 712899 712935 713079 713080 713112 713124 713126 713144 713149 713163 713171 713204 713209 713214 713218 713233 713236 713241 713270 713273 713300 713337 713352 713378 713401 713426 713449 713486 713491 713505 713548 713551 713620 713630 713659 713699 713709 713711 713712 713715 713716 713735 713740 713750 713757 713759 713769 713771 713776 713780 713786 713804 713814 713824 713826 713831 713836 713844 713857 713866 713884 713888 713901 713907 713919 713979 714255 714257 714259 714280 714286 714302 714317 714344 714351 714357 714358 714359 714380 714404 714413 714417 714420 714443 714459 714513 714578 714579 714595 714613 714618 714626 714628 714648 714655 714672 714677 714680 714685 714692 714693 714694 714702 714710 714721 714726 714729 714737 715527 715578 715606 715608 715610 715644 715854 715892 715975 715977 715981 715982 716007 716030 716048 716086 716142 716156 716201 716224 716495 716497 716658 716663 716680 716704 716708 716717 716723 716732 716736 716745 716749 717037 717312 717314 717318 717349 717373 717374 717375 717453 717660 717775 717791 717821 717937 718264 718275 718321 718331 718344 718354 718360 718362 718364 718372 718380 718381 718391 718395 718400 718408 718409 718416 718418 718437 718439 718440 718537 718546 718556 718586 718591 718596 718623 718632 718667 718676 718680 718692 718693 718713 718716 718720 718729 718737 718740 718765 718767 718786 718790 718800 718822 718826 718839 718880 718897 718900 718956 718979 719018 719019 719063 719084 719092 719100 719103 719118 719120 719492 719672 719749 719752 720024 720240 720374 720385 720423 720496 720531 720606 720612 720646 720671 720679 720680 720684 720686 720713 720731 720732 720736 720740 720749 720754 720771 720820 720822 720825 720830 720839 720845 720848 720851 720866 720870 720887 720895 720950 720989 721019 721020 721023 721026 721037 721039 721047 721052 721064 721069 721070 721074 721076 721090 721093 721098 721154 721161 721171 721346 721354 721556 721629 721711 721849 722061 722166 722191 722238 722258 722320 722321 722361 722394 722469 722512 722555 722557 722570 722684 722727 722738 722768 722818 722875 722883 722897 722908 722910 722924 722941 722959 722967 723021 723024 723044 723050 723129 723137 723148 723397 723422 723483 723517 723551 723611 723687 723723 723726 723763 723824 723848 723851 724077 725326 725929 725968 725980 725982 725983 725996 726017 726018 726019 726023 726039 726096 726135 726224 726226 726229 726231 726236 726237 726280 726299 726305 726315 726326 726359 726579 726582 726661 726671 726697 726769 726790 726795 726802 726803 726864 726867 726907 726939 726954 726990 727032 727100 727233 727236 727271 727385 727407 727452 727494 727496 727578 727582 727642 727671 727853 727855 727978

Distribution

US (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign) to countries of: Guam and Northern Mariana Islands.

Quantity

156,329 units (59,937 US, 96,792 OUS) in total