Hemotherm CE Dual Reservoir Cooler/Heater, Model No. 400CE Product Usage: The HEMOTHERM¿ Model 400CE Dual Reservoir Cooler/Heater is used to lower, maintain, or raise the temperature of the water flowing through a Blood Oxygenator / Heat Exchanger that is used to cool or warm blood during cardiopulmonary bypass procedures lasting six hours or less. The Hemotherm Dual Reservoir Cooler/Heater may also be used with a hyper / hypothermia blanket under the patient to provide warming through conductive heat transfer.

Recall: Z-0825-2019 · Reported February 20, 2019

Enforcement

Recall Number
Z-0825-2019
Event ID
81961
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cincinnati Sub-Zero Products LLC, a Gentherm Company
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 20, 2019
Initiation Date
December 19, 2018
Classification Date
February 13, 2019
Termination Date
April 24, 2020
Address
12011 Mosteller Rd, N/A, Cincinnati, OH, 45241-1528, United States

Description

Hemotherm CE Dual Reservoir Cooler/Heater, Model No. 400CE Product Usage: The HEMOTHERM¿ Model 400CE Dual Reservoir Cooler/Heater is used to lower, maintain, or raise the temperature of the water flowing through a Blood Oxygenator / Heat Exchanger that is used to cool or warm blood during cardiopulmonary bypass procedures lasting six hours or less. The Hemotherm Dual Reservoir Cooler/Heater may also be used with a hyper / hypothermia blanket under the patient to provide warming through conductive heat transfer.

Reason

Evidence supports that when customers performed a fuse replacement as part of a March 2017 field action, the fuse clips became stretched and were no longer mechanically tight with the fuse, creating a host spot that led to discoloration on the board due to overheating and shut down of the device.

Code Info

Serial numbers 151-10283CE through 171-10721CE

Distribution

Worldwide Distribution - US Nationwide in the states of: AL, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, and WI. The products were distributed to the following foreign countries: Columbia and Taiwan.

Quantity

396