FDA Enforcement
Class II
Terminated
ST AIA-PACK¿ PROG II, Progesterone II Assay, Part Number: 025239
Recall: Z-0825-2018
·
Reported March 7, 2018
Enforcement
- Recall Number
- Z-0825-2018
- Event ID
- 79176
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Tosoh Bioscience Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- March 7, 2018
- Initiation Date
- February 13, 2018
- Classification Date
- March 1, 2018
- Termination Date
- October 29, 2020
- Address
- 3600 Gantz Rd, N/A, Grove City, OH, 43123-1895, United States
Description
ST AIA-PACK¿ PROG II, Progesterone II Assay, Part Number: 025239
Reason
Falsely elevated progesterone assay values will occur when using certain progesterone assay part numbers due to cross reactivity of DHEA-S, a metabolite of the DHEA (dehydroepiandrosterone) supplement.
Code Info
All lots
Distribution
Distributed in 39 states: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
Quantity
5579