BCS XP Automated Blood Coagulation Analyzer System, Multipurpose system for in vitro coagulation studies, Catalog Numbers: 10459330 (BCS XP SYSTEM, COMPLETE), 10462449 (BCS XP ANALYZER SYSTEM), 10461894 (BCS XP), 10470625 (BCS XP REFURBISHED); IVD --- Device Listing D060801
Enforcement
- Recall Number
- Z-0825-2016
- Event ID
- 73099
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 2, 2016
- Initiation Date
- January 14, 2016
- Classification Date
- February 19, 2016
- Termination Date
- March 16, 2017
- Address
- 511 Benedict Ave, N/A, Tarrytown, NY, 10591-5005, United States
Description
BCS XP Automated Blood Coagulation Analyzer System, Multipurpose system for in vitro coagulation studies, Catalog Numbers: 10459330 (BCS XP SYSTEM, COMPLETE), 10462449 (BCS XP ANALYZER SYSTEM), 10461894 (BCS XP), 10470625 (BCS XP REFURBISHED); IVD --- Device Listing D060801
False short clotting times for PT on BCS and BCS XP with Dade Innovin can occur. This has also an effect for false low PT INR values. These false short clotting times were observed only with turbid, hemolytic and/or icteric samples from intensive care patients. The corresponding reaction curves of these samples showed a pre-peak, which can lead to falsely short clotting time. The result was not flagged by the existing check algorithms.
All serial numbers
Worldwide Distribution-US (nationwide) and the countries of Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bosnia Herzegovina, Brazil, Canada, Chile, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Hungary, Italy, Kuwait, Latvia, Lebanon, Libya, Macedonia, Mexico, Montenegro, Netherlands, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and Uruguay.
1895 units total