Cytocell 8 Square Template Slides (Glass)- microscope slides Reference: PCN008
Recall: Z-0824-2022
·
Reported March 30, 2022
Enforcement
- Recall Number
- Z-0824-2022
- Event ID
- 89747
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Cytocell Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- March 30, 2022
- Initiation Date
- February 14, 2022
- Classification Date
- March 24, 2022
- Termination Date
- June 12, 2023
- Address
- 418 Cambridge Science Park, Milton Road, Oxford Gene Technology, Cambridge, N/A, N/A, United Kingdom
Description
Cytocell 8 Square Template Slides (Glass)- microscope slides Reference: PCN008
Reason
Labelled with an incorrect expiry date on the outer label with an extended expiry date than the validated shelf life. The correct expiry date for the device is 2022-02 and not 2023-05 as shown on the outer label.
Code Info
Lot 210511-018
Distribution
US Nationwide distribution in the states of CT, TX.
Quantity
5 kits