FDA Enforcement Class II Ongoing

Custom procedural convenience kits and trays, gastroenterological & urological, labeled as: a) LAP KIDNEY MODULE, kit number AGKD50L; b) ROBOTIC PROSTATE PACK - 264852, kit number ANPR34S; c) LAP CHOLE - 213388, kit number BFLC02AD; d) GENERAL LAPAROSCOPY PACK, kit number GUGE09I; e) LAP CHOLE PACK, kit number HNLC80I; f) Lap Chole Ovarian Appy, kit number JSLC06I; g) LAP CHOLE, ST LUKES, kit number LMLP08Z; h) ROBOTIC PACK - 242561, kit number MHRB88AJ; i) KIT, GENERAL LAPAROSCOPY, kit number MMLP40F; j) KIT, MAJOR LAPAROTOMY, kit number MMML26F; k) LAP CHOLE - LAP HERNIA PACK - 282807, kit number OWLC07K; l) GENERAL LAPAROSCOPY SAH, kit number SAGL57H; m) ROBOTIC PACK SAH, kit number SARB30K; n) ROBOTIC PACK SAH, kit number SARB30L; o) ROBOTIC PACK SAH, kit number SARB30M; p) LAP CHOLE PACK, kit number SMLC53; q) ROBOTICS PACK, kit number SNRP85D; r) ROBOTIC PACK, kit number SSRO22C; s) ROBOTIC PACK, kit number SSRO22D; t) TPK DAVINCI PROSTATE SYSTEM, kit number TPDV16; u) ASC LAP CHOLE PACK (PS 42279), kit number UICH53Y; v) GENERAL LAPAROSCOPIC PACK, kit number WVGL60E

Recall: Z-0822-2024 · Reported February 7, 2024

Enforcement

Recall Number
Z-0822-2024
Event ID
93656
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
American Contract Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 7, 2024
Initiation Date
November 29, 2023
Classification Date
January 26, 2024
Address
7802 E Telecom Pkwy, Temple Terrace, FL, 33637-0928, United States

Description

Custom procedural convenience kits and trays, gastroenterological & urological, labeled as: a) LAP KIDNEY MODULE, kit number AGKD50L; b) ROBOTIC PROSTATE PACK - 264852, kit number ANPR34S; c) LAP CHOLE - 213388, kit number BFLC02AD; d) GENERAL LAPAROSCOPY PACK, kit number GUGE09I; e) LAP CHOLE PACK, kit number HNLC80I; f) Lap Chole Ovarian Appy, kit number JSLC06I; g) LAP CHOLE, ST LUKES, kit number LMLP08Z; h) ROBOTIC PACK - 242561, kit number MHRB88AJ; i) KIT, GENERAL LAPAROSCOPY, kit number MMLP40F; j) KIT, MAJOR LAPAROTOMY, kit number MMML26F; k) LAP CHOLE - LAP HERNIA PACK - 282807, kit number OWLC07K; l) GENERAL LAPAROSCOPY SAH, kit number SAGL57H; m) ROBOTIC PACK SAH, kit number SARB30K; n) ROBOTIC PACK SAH, kit number SARB30L; o) ROBOTIC PACK SAH, kit number SARB30M; p) LAP CHOLE PACK, kit number SMLC53; q) ROBOTICS PACK, kit number SNRP85D; r) ROBOTIC PACK, kit number SSRO22C; s) ROBOTIC PACK, kit number SSRO22D; t) TPK DAVINCI PROSTATE SYSTEM, kit number TPDV16; u) ASC LAP CHOLE PACK (PS 42279), kit number UICH53Y; v) GENERAL LAPAROSCOPIC PACK, kit number WVGL60E

Reason

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Code Info

a) AGKD50L, UDI/DI 00191072141156, Lot Numbers: 905231, exp. 4/5/2024; 638221, exp. 12/28/2023; 689221, exp. 5/21/2023; b) ANPR34S, UDI/DI 00191072171351, Lot Numbers: 721221, exp. 10/6/2023; 879231, exp. 5/1/2024; 896231, exp. 4/14/2024; 927231, exp. 3/14/2024; 956231, exp. 2/13/2024; 645221, exp. 12/21/2023; c) BFLC02AD, UDI/DI 00191072169075, Lot Numbers: 654221, exp. 12/12/2023; 647221, exp. 12/19/2023; 987231, exp. 1/13/2024; UDI/DI 00191072188830, Lot Numbers: 946231, exp. 2/23/2024; 896231, exp. 4/14/2024; d) GUGE09I, UDI/DI none, Lot Numbers: 768221, exp. 8/20/2023; e) HNLC80I, UDI/DI 00191072146267, Lot Numbers: 797221, exp. 7/22/2023; 679221, exp. 11/17/2023; 995231, exp. 1/25/2024; 911231, exp. 3/30/2024; f) JSLC06I, UDI/DI 00191072126061, Lot Numbers: 839221, exp. 6/10/2023; 766221, exp. 8/22/2023; g) LMLP08Z, UDI/DI none, Lot Numbers: 661221, exp. 12/5/2023; 987231, exp. 1/13/2024; 942231, exp. 2/27/2024; 893231, exp. 4/17/2024; h) MHRB88AJ, UDI/DI none, Lot Numbers: 735221, exp. 9/22/2023; 710221, exp. 10/17/2023; 997231, exp. 1/3/2024; 959231, exp. 2/10/2024; 904231, exp. 4/6/2024; i) MMLP40F, UDI/DI none, Lot Numbers: 933231, exp. 3/8/2024; 879231, exp. 5/1/2024; j) MMML26F, UDI/DI none, Lot Numbers: 989231, exp. 1/11/2024; 896231, exp. 4/14/2024; k) OWLC07K, UDI/DI none, Lot Numbers: 682221, exp. 11/14/2023; 646221, exp. 12/20/2023; 982231, exp. 1/18/2024; 928231, exp. 3/13/2024; l) SAGL57H, UDI/DI none, Lot Numbers: 687221, exp. 11/9/2023; 790221, exp. 7/29/2023; m) SARB30K, UDI/DI none, Lot Numbers: 845221, exp. 6/4/2023; 888221, exp. 4/22/2023; n) SARB30L, UDI/DI none, Lot Numbers: 752221, exp. 9/5/2023; o) SARB30M, UDI/DI none, Lot Numbers: 658221, exp. 12/8/2023; 700221, exp. 10/27/2023; 792221, exp. 7/27/2023; 949231, exp. 2/20/2024; 983231, exp. 1/17/2024; p) SMLC53, UDI/DI 00191072188465, Lot Numbers: 969231, exp. 1/31/2024; q) SNRP85D, UDI/DI none, Lot Numbers: 666221, exp. 11/30/2023; 720221, exp. 10/7/2023; 793221, exp. 7/26/2023; 853221, exp. 5/27/2023; 965231, exp. 2/4/2024; r) SSRO22C, UDI/DI none, Lot Numbers: 891221, exp. 04/19/2023; s) SSRO22D, UDI/DI none, Lot Numbers: 660221, exp. 5/31/2023; 698221, exp. 10/29/2023; 756221, exp. 9/1/2023; 804221, exp. 7/15/2023; 840221, exp. 6/9/2023; 933231, exp. 8/31/2023; 987231, exp. 1/13/2024; t) TPDV16, UDI/DI 00191072128904, Lot Numbers: 856221, exp. 5/24/2023; 744221, exp. 9/13/2023; 675221, exp. 11/21/2023; 932231, exp. 3/9/2024; u) UICH53Y, UDI/DI none, Lot Numbers: 897231, exp. 4/13/2024; 882231, exp. 4/28/2024; v) WVGL60E, UDI/DI 00191072144034, Lot Numbers: 766221, exp. 8/22/2023; 694221, exp. 11/2/2023; 952231, exp. 2/17/2024;

Distribution

US Nationwide distribution.

Quantity

3196 units