FDA Enforcement Class II Terminated

Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3

Recall: Z-0822-2020 · Reported January 29, 2020

Enforcement

Recall Number
Z-0822-2020
Event ID
83990
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Isopure Corp
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
January 29, 2020
Initiation Date
April 19, 2019
Classification Date
January 18, 2020
Termination Date
May 29, 2020
Address
11851 Plantside Dr, Louisville, KY, 40299-6328, United States

Description

Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3

Reason

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Code Info

IsoMix Pro UDI: (01)00852456007085; ISOMIX Serial Number IMP-838703 with Pump SN 180210D18

Distribution

The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.

Quantity

1