Viterion TeleHealth Monitor Model Number V500; Product Codes 08832984 (US) and 02080468 (UK) - Physiological patient monitor (without arrhythmia detection or alarms). Viterion V100, V100-BGM and V500 Telehealth Monitors transfer medical information.

Recall: Z-0822-2013 · Reported February 20, 2013

Enforcement

Recall Number
Z-0822-2013
Event ID
64111
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Viterion TeleHealthcare Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
February 20, 2013
Initiation Date
November 15, 2012
Classification Date
February 14, 2013
Termination Date
February 16, 2017
Address
555 White Plains Rd, Tarrytown, NY, 10591-5101, United States

Description

Viterion TeleHealth Monitor Model Number V500; Product Codes 08832984 (US) and 02080468 (UK) - Physiological patient monitor (without arrhythmia detection or alarms). Viterion V100, V100-BGM and V500 Telehealth Monitors transfer medical information.

Reason

Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (below the meter's measurable range) blood glucose readings from Bayer's Contour Elite XL blood glucose meters, or from Roche's Accu-Chek Advantage 03 and 04 meters to healthcare professionals.

Code Info

All units are affected. Serial Numbers A3303HA000201, A3303HA000202, A3303HA000203, A3303HA000204, A3303HA000205, A3304HA000206, A3304HA000207, A3304HA000208, A3304HA000209, A3304HA000210, A3305HA000211, A3305HA000212, A3305HA000213, A3305HA000214, A3305HA000215, A3306HA000216, A3306HA000217, A3306HA000218, A3306HA000219, A3306HA000220, A3310HA000221, A3310HA000222, A3310HA000223, A3310HA000224, A3310HA000225, A3404HA000226, A3404HA000227, A3404HA000228, A3404HA000229, A3404HA000230, A3404HA000231, A3404HA000232, A3404HA000233, A3404HA000234, A3404HA000235, A3404HA000236, A3404HA000237, A3404HA000238, A3404HA000239, A3404HA000240, A3404HA000241, A3404HA000242, A3404HA000243, A3404HA000244, A3404HA000245, A3404HA000246, A3404HA000247, A3404HA000248, A3404HA000249, A3404HA000250, A3602HA000252, A3602HA000253, A3602HA000254, A3602HA000255, A3602HA000256, A3602HA000257, A3602HA000258, A3602HA000259, A3602HA000260, A3602HA000261, A3602HA000262, A3602HA000263, A3602HA000264, A3602HA000265, A3602HA000266, A3701HA000267, A3701HA000268, A3701HA000269, A3701HA000270, A3701HA000271, A3701HA000272, A3701HA000273, A3701HA000274, A3701HA000275, A3701HA000276, A3701HA000277, A3701HA000278, A3701HA000279, A3701HA000280, A3701HA000281, A3701HA000282, A3701HA000283, A3701HA000284, A3701HA000285, A3701HA000286, A3701HA000287, A3701HA000288, A3701HA000289, A3701HA000290, A3701HA000291, A3818HA000292, A3818HA000293, A3818HA000294, A3818HA000295, A3818HA000296, A3818HA000297, A3818HA000298, A3818HA000299, A3818HA000300, A3818HA000301, A3818HA000302, A3818HA000303, A3818HA000304, A3818HA000305, A3818HA000306, A3818HA000307, A3818HA000308, A3818HA000309, A3818HA000310, A3818HA000311, A3818HA000312, A3818HA000313, A3818HA000314, A3818HA000315, A3818HA000316, A3905HA000317, A3905HA000318, A3905HA000319, A3905HA000320, A3905HA000321, A3905HA000322, A3905HA000323, A3905HA000324, A3905HA000325, A3905HA000326, A3905HA000327, A3905HA000328, A3905HA000329, A3905HA000330, A3905HA000331, A3905HA000332, A3905HA000333, A3905HA000334, A3905HA000335, A3905HA000336, A3905HA000337, A3905HA000338, A3905HA000339, A3905HA000340, A3905HA000341, A3D01HA000342, A3D01HA000343, A3D01HA000344, A3D01HA000345, A3D01HA000346, A3D01HA000347, A3D01HA000348, A3D01HA000349, A3D01HA000350, A3D01HA000351, A3D01HA000352, A3D01HA000353, A3D01HA000354, A3D01HA000355, A3D01HA000356, A3D01HA000357, A3D01HA000358, A3D01HA000359, A3D01HA000360, A3D01HA000361, A3D01HA000362, A3D01HA000363, A3D01HA000364, A3D01HA000365, A3D01HA000366, A3D01HA000367, A3D01HA000368, A3D01HA000369, A3D01HA000370, A3D01HA000371, A3D01HA000372, A3D01HA000373, A3D01HA000374, A3D01HA000375, A3D01HA000376, A3D01HA000377, A3D01HA000378, A3D01HA000379, A3D01HA000380, A3D01HA000381, A3D01HA000382, A3D01HA000383, A3D01HA000384, A3D01HA000385, A3D01HA000386, A3D01HA000387, A3D01HA000388, A3D01HA000389, A3D01HA000390, A4113HA000391, A4113HA000392, A4113HA000393, A4113HA000394, A4113HA000395, A4113HA000396, A4113HA000397, A4113HA000398, A4113HA000399, A4113HA000400, A4113HA000401, A4113HA000402, A4113HA000403, A4113HA000404, A4113HA000405, A4113HA000406, A4113HA000407, A4113HA000408, A4113HA000409, A4113HA000410, A4113HA000411, A4113HA000412, A4113HA000413, A4113HA000414, A4113HA000415, A4113HA000416, A4113HA000417, A4113HA000418, A4113HA000419, A4113HA000420, A4113HA000421, A4113HA000422, A4113HA000423, A4113HA000424, A4113HA000425, A4113HA000426, A4113HA000427, A4113HA000428, A4113HA000429, A4113HA000430, A4113HA000431, A4113HA000432, A4113HA000433, A4113HA000434, A4113HA000435, A4113HA000436, A4113HA000437, A4113HA000438, A4113HA000439, A4113HA000440, A4113HA000441, A4526MA000442, A4526MA000443, A4526MA000444, A4526MA000446, A4712MA000445, A4712MA000447, A4712MA000448, A4712MA000449, A4712MA000450, A4712MA000451, A4712MA000452, A4712MA000453, A4712MA000454, A4712MA000455, A4712MA000456, A4712MA000457, A4712MA000458, A4712MA000459, A4712MA000460, A4712MA000461, A4712MA000462, A4712MA000463, A4712MA000464, A4712MA000465, A4712MA000466, A4712MA000467, A4712MA000468, A4712MA000469, A4712MA000470, A4712MA000471, A4712MA000472, A4712MA000473, A4712MA000474, A4712MA000475, A4712MA000476, A4712MA000477, A4712MA000478, A4712MA000479, A4712MA000480, A4712MA000481, A4712MA000482, A4712MA000483, A4712MA000484, A4712MA000485, A4712MA000486, A4712MA000487, A4712MA000488, A4712MA000489, A4712MA000490, A4712MA000491, A4712MA000492, A4712MA000493, A4712MA000494, A4712MA000495, A4712MA000496, A4712MA000497, A4712MA000498, A4712MA000499, A4712MA000500, A4712MA000501, A4712MA000502, A4712MA000503, A4712MA000504, A4712MA000505, A4712MA000506, A5225MA000507, A5225MA000508, A5225MA000509, A5225MA000510, A5225MA000511, A5225MA000512, A5225MA000513, A5225MA000514, A5225MA000515, A5225MA000516, A5225MA000517, A5225MA000518, A5225MA000519, A5225MA000520, A5225MA000521, A5225MA000522, A5225MA000523, A5225MA000524, A5225MA000525, A5225MA000526, A5225MA000527, A5225MA000528, A5225MA000529, A5225MA000530, A5225MA000531, A5225MA000532, A5225MA000533, A5225MA000534, A5225MA000535, A5225MA000536, A5225MA000537, A5225MA000538, A5225MA000539, A5225MA000540, A5225MA000541, A5225MA000542, A5225MA000543, A5225MA000544, A5225MA000545, A5225MA000546, A5225MA000547, A5225MA000548, A5225MA000549, A5225MA000550, A5225MA000551, A5225MA000552, A5225MA000553, A5225MA000554, A5225MA000555, A5225MA000556, A5225MA000557, A5225MA000558, A5225MA000559, A5225MA000560, A5225MA000561, A5225MA000562, A5225MA000563, A5225MA000564, A5225MA000565, A5225MA000566, A5225MA000567, A5225MA000568, A5225MA000569, A5225MA000570, A5225MA000571, A5225MA000572, A5225MA000573, A5225MA000574, A5225MA000575, A5225MA000576, A5225MA000577, A5225MA000578, A5225MA000579, A5225MA000580, A5225MA000581, A5225MA000582, A5225MA000583, A5225MA000584, A5225MA000585, A5225MA000586, A5225MA000587, A5225MA000588, A5225MA000589, A5225MA000590 A5225MA000591, A5225MA000592, A5225MA000593, A5225MA000594, A5225MA000595, A5225MA000596, A5225MA000597, A5225MA000598, A5225MA000599, A5225MA000600, A5225MA000601, A5225MA000602, A5225MA000603, A5225MA000604, A5225MA000605, A5225MA000606, A5225MA000607, A5225MA000608, A5225MA000609, A5225MA000610, A5225MA000611, A5225MA000612, A5225MA000613, A5225MA000614, A5225MA000615, A5225MA000616, A5225MA000617, A5225MA000618, A5225MA000619, A5225MA000620, A5225MA000621, A5225MA000622, A5225MA000623, A5225MA000624, A5225MA000625, A5225MA000626, A5225MA000627, A5225MA000628, A5225MA000629, A5225MA000630, A5225MA000631, A5225MA000632, A5225MA000633, A5225MA000634, A5225MA000635, A5225MA000636, A5225MA000637, A5225MA000638, A5225MA000639, A5225MA000640, A5225MA000641, A5225MA000642, A5225MA000643, A5225MA000644, A5225MA000645, A5225MA000646, A5225MA000647, A5225MA000648, A5225MA000649, A5225MA000650, A5225MA000651, A5225MA000652, A5225MA000653, A5225MA000654, A5225MA000655, A5225MA000656, A5225MA000657, A5225MA000658, A5225MA000659, A5225MA000660, A5225MA000661, A5225MA000662, A5225MA000663, A5225MA000664, A5225MA000665, A5225MA000666, A5225MA000667, A5225MA000668, A5225MA000669, A5225MA000670, A5225MA000671, A5225MA000672, A5225MA000673, A5225MA000674, A5225MA000675, A5225MA000676, A5225MA000677, A5225MA000678, A5225MA000679, A5225MA000680, A5225MA000681, A5225MA000682, A5225MA000683, A5225MA000684, A5225MA000685, A5225MA000686, A5225MA000687, A5225MA000688, A5225MA000689, A5225MA000690, A5225MA000691, A5225MA000692, A5225MA000693, A5225MA000694, A5225MA000695, A5225MA000696, A5225MA000697, A5225MA000698, A5225MA000699, A5225MA000700, A5225MA000701, A5225MA000702, A5225MA000703, A5225MA000704, A5225MA000705, A5225MA000706, A5225MA000707, A5225MA000708, A5225MA000709, A5225MA000710, A5225MA000711, A5225MA000712, A5225MA000713, A5225MA000714, A5225MA000715, A5225MA000716, A5225MA000717, A5225MA000718, A5225MA000719, A5225MA000720, A5225MA000721, A5225MA000722, A5225MA000723, A5225MA000724, A5225MA000725, A5225MA000726, A5225MA000727, A5225MA000728, A5225MA000729, A5225MA000730, A5225MA000731, A5225MA000732, A5225MA000733, A5225MA000734, A5311MA000735, A5311MA000736, A5314MA000737, A5314MA000738, A5314MA000739, A5314MA000740, A5314MA000741, A5314MA000742, A5314MA000743, A5314MA000744, A5314MA000745, A5314MA000746, A5314MA000747, A5314MA000748, A5314MA000749, A5314MA000750, A5314MA000751, A5314MA000752, A5314MA000753, A5314MA000754, A5314MA000755, A5314MA000756, A5314MA000757, A5314MA000758, A5314MA000759, A5314MA000760, A5314MA000761, A5414MA000762, A5414MA000763, A5414MA000764, A5414MA000765, A5414MA000766, A5414MA000767, A5414MA000768, A5414MA000769, A5414MA000770, A5414MA000771, A5414MA000772, A5414MA000773, A5414MA000774, A5414MA000775, A5414MA000776, A5414MA000777, A5414MA000778, A5414MA000779, A5414MA000780, A5414MA000781, A5414MA000782, A5414MA000783, A5414MA000784, A5414MA000785, A5414MA000786, A5414MA000787, A5414MA000788, A5414MA000789, A5414MA000790, A5414MA000791, A5414MA000792, A5414MA000793, A5414MA000794, A5414MA000795, A5414MA000796, A5414MA000797, A5414MA000798, A5414MA000799, A5414MA000800, A5414MA000801, A5421MA000802, A5421MA000803, A5421MA000804, A5421MA000805, A5421MA000806, A5421MA000807, A5421MA000808, A5421MA000809, A5421MA000810, A5516MA000811, A5516MA000812, A5516MA000813, A5516MA000814, A5516MA000815, A5516MA000816, A5516MA000817, A5516MA000818, A5516MA000819, A5516MA000820, A5527MA000821, A5527MA000822, A5527MA000823, A5527MA000824, A5527MA000825, A5527MA000826, A5527MA000827, A5527MA000828, A5527MA000829, A5527MA000830, A5527MA000831, A5527MA000832, A5527MA000833, A5527MA000834, A5527MA000835, A5527MA000836, A5527MA000837, A5527MA000838, A5527MA000839, A5527MA000840, A5527MA000841, A5527MA000842, A5527MA000843, A5527MA000844, A5527MA000845, A5527MA000846, A5527MA000847, A5527MA000848, A5527MA000849, A5527MA000850, A5527MA000851, A5527MA000852, A5527MA000853, A5527MA000854, A5527MA000855, A5527MA000856, A5527MA000857, A5527MA000858, A5527MA000859, A5527MA000860, A5D15MA000891, A5D15MA000892, A5D15MA000893, A5D15MA000894, A5D15MA000895, A5D15MA000896, A5D15MA000897, A5D15MA000898, A5D15MA000899, A5D15MA000900, A5D15MA000901, A5D15MA000902, A5D15MA000903, A5D15MA000904, A5D15MA000905, A5D15MA000906, A5D15MA000907, A5D15MA000908, A5D15MA000909, A5D15MA000910, A5D15MA000911, A5D15MA000912, A5D15MA000913, A5D15MA000914, A5D15MA000915, A5D15MA000916, A5D15MA000917, A5D15MA000918, A5D15MA000919, A5D15MA000920, A5N08MA000861, A5N08MA000862, A5N08MA000863, A5N08MA000864, A5N08MA000865, A5N08MA000866, A5N08MA000867, A5N08MA000868, A5N08MA000869, A5N08MA000870, A5N08MA000871, A5N08MA000872, A5N08MA000873, A5N08MA000874, A5N08MA000875, A5N08MA000876, A5N08MA000877, A5N08MA000878, A5N08MA000879, A5N08MA000880, A5N08MA000881, A5N08MA000882, A5N08MA000883, A5N08MA000884, A5N08MA000885, A5N08MA000886, A5N08MA000887, A5N08MA000888, A5N08MA000889, A5N08MA000890, A6112MA000921, A6112MA000922, A6112MA000923, A6112MA000924, A6112MA000925, A6112MA000926, A6112MA000927, A6112MA000928, A6112MA000929, A6112MA000930, A6112MA000931, A6112MA000932, A6112MA000933, A6112MA000934, A6112MA000935, A6112MA000936, A6112MA000937, A6112MA000938, A6112MA000939, A6112MA000940, A6112MA000941, A6112MA000942, A6112MA000943, A6112MA000944, A6112MA000945, A6112MA000946, A6112MA000947, A6112MA000948, A6112MA000949, A6112MA000950, A6824MA000951, A6824MA000952, A6824MA000953, A6824MA000954, A6824MA000955, A6824MA000956, A6824MA000957, A6824MA000958, A6824MA000959, A6824MA000960

Distribution

Worldwide Distribution-USA (nationwide) and the United Kingdom.

Quantity

Domestic - 296 units, Foreign - 73 units