FDA Enforcement Class II Terminated

OTTO BOCK Kenevo knee joints Model 3C60=ST

Recall: Z-0815-2018 · Reported March 7, 2018

Enforcement

Recall Number
Z-0815-2018
Event ID
79019
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Otto Bock Healthcare Product
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 7, 2018
Initiation Date
September 12, 2017
Classification Date
March 1, 2018
Termination Date
January 19, 2021
Address
Brehmstrase 16, N/A, Vienna, N/A, N/A, Austria

Description

OTTO BOCK Kenevo knee joints Model 3C60=ST

Reason

Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used longer than 1.45 million cycles (which represents approximately 2.3 years of use of the knee joint under heavy usage conditions), AND the pylon clamp is overtightened according to the labeled torque specification, AND the patient weighs at most 125kg it may be remote that the pylon breaks, and causing the patient to fall. There have been no failures or complaints reported directly attributed to the failure mode collapsed pylon. In the US Market and in Canada, labeling requires a condition based service at three years. In the rest of the world, a mandatory two year service interval is specified. Only the devices in distribution in the United States and Canada are affected by this action.

Code Info

Serial Numbers: 201520028, 201520027, 20150026, and 201530040,

Distribution

AK, AR, AZ, CA, CT, DC, GA, IL, IN, KY, LA, MD, MI, MN, NC, ND, NJ, MN, NY, OH, OK, OR, PA, PR, TX, VA, and WA AU, BE, DE, FR, GB, IT, KW, LU, NL, NO, AT, SA, SE, CH, ES, ZA, and HU

Quantity

4