SmartPath to dStream for 1.5T, Model No. 781260, 782112
Enforcement
- Recall Number
- Z-0812-2024
- Event ID
- 93788
- Classification
- Class III
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Philips North America
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 31, 2024
- Initiation Date
- December 29, 2023
- Classification Date
- January 25, 2024
- Address
- 222 Jacobs St, N/A, Cambridge, MA, 02141-2296, United States
Description
SmartPath to dStream for 1.5T, Model No. 781260, 782112
An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.
Model No. 781260 UDI-DI 00884838095076 Serial Numbers 21020 9024 8657 21200 21223 18948 32131; Model No. 782112 UDI-DI 00884838098886 Serial Numbers 30089 20204
Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.
7 US; 16 ROW total