FDA Enforcement Class II Terminated

Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.

Recall: Z-0810-2013 · Reported February 20, 2013

Enforcement

Recall Number
Z-0810-2013
Event ID
63991
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Mizuho OSI
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 20, 2013
Initiation Date
December 17, 2012
Classification Date
February 13, 2013
Termination Date
April 15, 2016
Address
30031 Ahern Ave, Union City, CA, 94587-1234, United States

Description

Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.

Reason

Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.

Code Info

Table model 5855, sub assembly 5855-901 Orthopedic Table Spar, manufactured between 1996 and 2007.

Distribution

USA Distribution -- Worldwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and WY., and the countries of Australia, Canada, Egypt, Finland, France, Germany, China, Israel, Italy, Japan, Netherlands, New Zealand, P.R. China, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Taiwan R.O.C., UAE and UK.

Quantity

869 devices