FDA Enforcement Class II Ongoing

Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.

Recall: Z-0808-2025 · Reported January 15, 2025

Enforcement

Recall Number
Z-0808-2025
Event ID
95832
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Aerin Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 15, 2025
Initiation Date
November 12, 2024
Classification Date
January 6, 2025
Address
232 E Caribbean Dr, Sunnyvale, CA, 94089-1007, United States

Description

Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.

Reason

Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console.

Code Info

Lot #10235, exp. 2026-05-01; UDI (01)08886479300269(17)260501(10)10235

Distribution

US Nationwide Distribution: AL, AZ, CA, CO, GA, ID, IL, IN, KY LA, MI, MN, NY, NE, NY, OH, OR PA, SC, TN, TX, UT, VA, WA, WI, and WV.

Quantity

479 units