FDA Enforcement Class II Ongoing

GE Healthcare CARESCAPE Central Station (CSCS) V2.0x MAI700 (Integrated), REF 2082278-001, and MAS700 (Desktop, REF 2082279-001.

Recall: Z-0808-2022 · Reported March 23, 2022

Enforcement

Recall Number
Z-0808-2022
Event ID
89576
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
GE Healthcare, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 23, 2022
Initiation Date
January 28, 2022
Classification Date
March 16, 2022
Address
3000 N Grandview Blvd, N/A, Waukesha, WI, 53188-1615, United States

Description

GE Healthcare CARESCAPE Central Station (CSCS) V2.0x MAI700 (Integrated), REF 2082278-001, and MAS700 (Desktop, REF 2082279-001.

Reason

If the CARESCAPE Central Station v2.0 is used with an unapproved keyboard, the audio can be muted resulting in loss of audible alarms.

Code Info

Version 2.0.x (v2.0.0, v2.0.1, v2.0.2, or v2.0.3) with serial numbers beginning with SKN for the MAI700 Integrated, UDI 00840682109666 and serial numbers beginning with SNF for the MAI700 Integrated, UDI 0100840682109604; or serial numbers beginning with SNF for the MAS700 Desktop, UDI 00840682109604.

Distribution

Worldwide distribution - US nationwide, including Puerto Rico. There was also government/military distribution. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cote d'lvoire, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, Germany, Ghana, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Republic of Korea, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Myanmar, Netherlands, Nigeria, Norway, Pakistan, Palestine, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.

Quantity

10,445 units