100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002-000ENG) 230V NEPTUNE 3 ROVER SPANISH (0703-002-000ES)

Recall: Z-0807-2025 · Reported January 15, 2025

Enforcement

Recall Number
Z-0807-2025
Event ID
95856
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 15, 2025
Initiation Date
December 3, 2024
Classification Date
January 6, 2025
Address
1941 Stryker Way, Portage, MI, 49002-9711, United States

Description

100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002-000ENG) 230V NEPTUNE 3 ROVER SPANISH (0703-002-000ES)

Reason

Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of suction during a procedure.

Code Info

0703 002 000ENG GTIN 07613327382082, Lot Numbers; 2427418723 2427423863 2427418453 2427423873 2427418743 2427419773; 0703 002 000ES GTIN 07613327382068, Lot Numbers: 2427418453; 0703 003 000 GTIN 07613327381948, Lot Numbers: 2427418093 2427418053 2427418073 2427418233 2427418013 2427418723; 0711 001 000 GTIN 07613327576887, Lot Numbers: 2429720623 2429721033 2429720903 2429721203 2429720973 2429721223 2429721023 2303501403 2323610223 2403600253;

Distribution

US (Domestic) distribution to states of: WA, GA, and IL. OUS (International) distribution to Japan and Netherlands.

Quantity

14 (Expanded to 21 units on 10/9/25)