FDA Enforcement Class III Ongoing

Achieva 1.5T, Model No. 781178, 781196, 781296

Recall: Z-0807-2024 · Reported January 31, 2024

Enforcement

Recall Number
Z-0807-2024
Event ID
93788
Classification
Class III
Status
Ongoing
Product Type
Devices
Firm
Philips North America
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 31, 2024
Initiation Date
December 29, 2023
Classification Date
January 25, 2024
Address
222 Jacobs St, N/A, Cambridge, MA, 02141-2296, United States

Description

Achieva 1.5T, Model No. 781178, 781196, 781296

Reason

An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.

Code Info

Model No. 781178 UDI-DI N/A Serial Number 20150; Model No. 781196 UDI-DI N/A Serial Nos. 22270 22095 21113; Model No. 781296 UDI-DI 00884838004108 Serial Nos. 33406 33547 32773 32789 32388

Distribution

Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.

Quantity

7 US; 16 ROW total