FDA Enforcement Class II Ongoing

Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972

Recall: Z-0803-2025 · Reported January 15, 2025

Enforcement

Recall Number
Z-0803-2025
Event ID
95871
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 15, 2025
Initiation Date
November 26, 2024
Classification Date
January 3, 2025
Address
1000 Lake Hazeltine Dr, Chaska, MN, 55318-1037, United States

Description

Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972

Reason

Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.

Code Info

UDI: (01)15099590201937(17)241130(11)231201; Lot 339071

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Quantity

362 units