FDA Enforcement Class II Terminated

VITROS Automation Solutions, VAS, Catalog Numbers 6844300 & 6844301 - Product Usage: intended to automate pre-analytical and post-analytical sample processing in the clinical laboratory.

Recall: Z-0801-2021 · Reported January 13, 2021

Enforcement

Recall Number
Z-0801-2021
Event ID
86972
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 13, 2021
Initiation Date
November 20, 2020
Classification Date
January 7, 2021
Termination Date
October 24, 2023
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101, United States

Description

VITROS Automation Solutions, VAS, Catalog Numbers 6844300 & 6844301 - Product Usage: intended to automate pre-analytical and post-analytical sample processing in the clinical laboratory.

Reason

A software anomaly may cause an aliquoted sample to be labelled as the incorrect sample. This may lead to results being associated with the wrong patient sample and potentially lead to inappropriate intervention with the potential for injury to the patient.

Code Info

enGen Laboratory Automation Systems and VITROS Automation Solutions Using All TCAutomationTM configured with Thermo Scientific Aliquoter & Labeller Module and using Software Versions 4.6 and below

Distribution

Worldwide distribution - US Nationwide distribution including the states of CA, CO, MA, MD, NC, OH, TX, VA, VT, and Belgium, Bermuda, Canada, China, France, Italy, Japan, Mexico, Norway, Portugal, Singapore, Spain, Sweden, and United Kingdom.

Quantity

40 units