Computed Tomography X-Ray System (Brilliance CT Big Bore Oncology & Brilliance CT Big Bore Radiology) Distributed by Philips Healthcare System, Cleveland, OH. Produces cross-sectional body images by computer reconstruction of x-ray transmission data taken at different angles and planes.
Enforcement
- Recall Number
- Z-0801-2014
- Event ID
- 67168
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Medical Systems (Cleveland) Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 29, 2014
- Initiation Date
- December 4, 2013
- Classification Date
- January 21, 2014
- Termination Date
- July 1, 2015
- Address
- 595 Miner Rd, Cleveland, OH, 44143-2131, United States
Description
Computed Tomography X-Ray System (Brilliance CT Big Bore Oncology & Brilliance CT Big Bore Radiology) Distributed by Philips Healthcare System, Cleveland, OH. Produces cross-sectional body images by computer reconstruction of x-ray transmission data taken at different angles and planes.
Machine Name in TumorLOC Basic Mode is obscured. When TumorLOC is in Basic Mode and Set Vertical Laser equals Yes, the Machine Name label and field are covered by the controls above it. The Machine Name cannot be viewed or changed.
Model #: 728243; 7133, 7170, 7225, 7262, 7265, 7701, 7706, 7779, 7788, 7792, 7797, 7802, 7805, 7806, 7807, 7808, 7810, 7811, 7812, 7814, 7815, 7816, 7824, 7831, 7832, 7833, 7834 ,7835. Model #728244; 7621, 7636, 7809 & 7844.
Worldwide Distribution-USA (nationwide) including the states of AK, MD, NH, NV, OH, and OR, and the countries of Australia, China, France, India, Indonesia, Italy, Korea, Mexico, Poland, Romania, Russia, Spain, Switzerland, Taiwan, Thailand & United Arab Emirates.
32 units