DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- Carestream Health, Inc. The DRX-Revolution Mobile X-Ray System is a mobile imaging system that incorporates a self contained x-ray generator, image receptor, imaging display and software for acquiring medical diagnostic images outside of a standard x-ray room. The DRX-Revolution Mobile X-Ray System has a motor for driving the system so that the operator does not have to push the system, but can walk behind the system while it is moving from one location to another.
Enforcement
- Recall Number
- Z-0800-2015
- Event ID
- 69826
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Carestream Health Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 24, 2014
- Initiation Date
- November 25, 2014
- Classification Date
- December 12, 2014
- Termination Date
- June 6, 2016
- Address
- 150 Verona St, N/A, Rochester, NY, 14608-1733, United States
Description
DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- Carestream Health, Inc. The DRX-Revolution Mobile X-Ray System is a mobile imaging system that incorporates a self contained x-ray generator, image receptor, imaging display and software for acquiring medical diagnostic images outside of a standard x-ray room. The DRX-Revolution Mobile X-Ray System has a motor for driving the system so that the operator does not have to push the system, but can walk behind the system while it is moving from one location to another.
While servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution Mobile X-Ray System in which bolts that secure the x-ray tube assembly/boom to the device mounting block may break. It has been determined that this issue is a result of an improperly assembled part which holds the x-ray
Serial Numbers 101 to 764
Worldwide Distribution: US (nationwide) and countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, ITALY, JAPAN, KOREA, KUWAIT, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, and UNITED KINGDOM.
629 units (Domestic: 358 units; Foreign: 271 units)