FDA Enforcement
Class II
Terminated
Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine.
Recall: Z-0800-2014
·
Reported January 29, 2014
Enforcement
- Recall Number
- Z-0800-2014
- Event ID
- 67030
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 29, 2014
- Initiation Date
- December 17, 2013
- Classification Date
- January 21, 2014
- Termination Date
- August 26, 2014
- Address
- 500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States
Description
Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine.
Reason
An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits, and it is possible for the user to misinterpret or edit the result.
Code Info
ADVIA 1200 with software versions: V2.00 and V2.01 ADVIA 1650 with software versions V3.52 and V4.01
Distribution
Worldwide Distribution. USA nationwide and the country of Canada.
Quantity
1563