FDA Enforcement Class II Terminated

Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: A screwdriver facilitates implant insertion and removal.

Recall: Z-0799-2020 · Reported January 22, 2020

Enforcement

Recall Number
Z-0799-2020
Event ID
84350
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Trilliant Surgical, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 22, 2020
Initiation Date
March 14, 2019
Classification Date
January 10, 2020
Termination Date
March 16, 2022
Address
727 N Shepherd Dr Ste 100, Houston, TX, 77007-1320, United States

Description

Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: A screwdriver facilitates implant insertion and removal.

Reason

A screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive feature causing the driver to be unable to retain screws as intended.

Code Info

Part Number: 320-35-003 Lot Number: TSL006617 UDI: (01)00812926026893(10)TSL006617

Distribution

US Nationwide distribution in the states of CO, CT, IA, TX and WI. OUS: None

Quantity

20 units