Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).

Recall: Z-0797-2016 · Reported February 17, 2016

Enforcement

Recall Number
Z-0797-2016
Event ID
73069
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Orthofix, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 17, 2016
Initiation Date
January 8, 2016
Classification Date
February 10, 2016
Termination Date
November 16, 2016
Address
3451 Plano Pkwy, N/A, Lewisville, TX, 75056-9453, United States

Description

Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).

Reason

There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.

Code Info

BS535418105, BS535419118, BS535420030, BS535419124, BS562325021, BS535418158, BS535418038, BS535418138, BS535418012, BS535419006, BS535420039, BS535418127, BS549024010, BS535419141, BS535418081, BS535418111, BS535418117, BS535420041, BS535418144, BS535420081, BS535419151, BS562325006, BS546825008, BS535418002, BS535418013, BS535418131, BS535420006, BS535420010, BS535420026, BS535420047, BS556219012, BS556219026, BS535418109, BS535420082, BS557717009, and BS562325023.

Distribution

Worldwide Distribution -- US, Spain, Italy, and Australia.

Quantity

36 units