OSCOR ADELANTE BREEZEWAY 8F C55¿ S79CM D84CM Catalog # AB081044 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.

Recall: Z-0794-2017 · Reported December 28, 2016

Enforcement

Recall Number
Z-0794-2017
Event ID
75785
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Oscor, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
December 28, 2016
Initiation Date
October 5, 2016
Classification Date
December 19, 2016
Termination Date
July 26, 2019
Address
3816 Desoto Blvd, N/A, Palm Harbor, FL, 34683-1618, United States

Description

OSCOR ADELANTE BREEZEWAY 8F C55¿ S79CM D84CM Catalog # AB081044 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.

Reason

Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.

Code Info

Lot #'s: C1-12152 Exp. November 30, 2018, C1-09518 Exp. December 21, 2017, C1-09748 Exp. February 16, 2018, C1-09899 Exp. March 12, 2018, C1-10040 Exp. April 5, 2018, C1-10040 Exp. April 5, 2018, C1-10149 Exp. April 23, 2018, C1-10330 Exp. May 11, 2018, C1-10329 Exp. May 11, 2018, C1-10382 Exp. May 17, 2018, C1-11571 Exp. September 25, 2018, C1-11571 Exp. September 25, 2018, and OR-04757 Exp. December 22, 2018

Distribution

Worldwide Distribution - US including CA, NJ, MD, TX, FL, NC, MO, and Puerto Rico. Internationally to AUSTRIA, BOLIVIA, CANADA, CHILE, CYPRUS, NETHERLANS, EGYPT, GERMANY, HONG KONG, ISRAEL, ITALY, SWITZERLAND, and VIETNAM

Quantity

991